BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM
Report
- Report Number
- 1423507-2023-00026
- Event Type
- Malfunction
- Date Received
- January 12, 2023
- Date of Event
- June 13, 2022
- Report Date
- January 12, 2023
- Manufacturer
- CAREFUSION, INC
- Product Code
- FQH
- UDI-DI
- 00382909101003
- PMA / PMN Number
- K202071
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRO CODES: FQH (LAVAGE, JET) AND FRO (DRESSING, WOUND, DRUG). NO PHOTOS OR PHYSICAL SAMPLES WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION; THEREFORE, THIS COMPLAINT COULD NOT BE VERIFIED. THE REPORTED LOT'S (1854590) DEVICE HISTORY RECORDS WERE REVIEWED. ALL PROCESS AND FINAL INSPECTIONS MEET SPECIFICATIONS. AN IN-DEPTH INVESTIGATION AS PART OF THAT CAPA HAS BEEN PERFORMED TO IDENTIFY THE CAUSE OF THE SURGIPHOR BOTTLE CRACKING DURING USE. AT THIS TIME THIS INVESTIGATION DISCOVERED THAT THE COMBINATION OF THE FOLLOWING VARIABLES ARE PROBABLE CAUSES THAT CONTRIBUTE TO THE FAILURE MODE OF SURGIPHOR CONTAINERS CRACKING DURING USE: AGING, BOTTLE WALL THICKNESS/BOTTLE WEIGHT, INSPECTED PRODUCT RETURNED INTO THE MANUFACTURING LOT, AND UPPER END OF THE GAMMA STERILIZATION DOSAGES OF THE PRODUCT. A FOLLOW-UP WILL BE SENT IF ADDITIONAL INFORMATION IS RECEIVED.
A BOTTLE BURST WHILE BEING SQUEEZED DURING APPLICATION OF PRODUCT ON PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1535057 | BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM | SURGIPHOR | FQH | CAREFUSION, INC | 910100 | 1854590 | 00382909101003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |