FDA Adverse Event Malfunction Summary report: N

BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM

MDR report key: 16147273 · Received January 12, 2023

Report

Report Number
1423507-2023-00026
Event Type
Malfunction
Date Received
January 12, 2023
Date of Event
June 13, 2022
Report Date
January 12, 2023
Manufacturer
CAREFUSION, INC
Product Code
FQH
UDI-DI
00382909101003
PMA / PMN Number
K202071
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRO CODES: FQH (LAVAGE, JET) AND FRO (DRESSING, WOUND, DRUG). NO PHOTOS OR PHYSICAL SAMPLES WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION; THEREFORE, THIS COMPLAINT COULD NOT BE VERIFIED. THE REPORTED LOT'S (1854590) DEVICE HISTORY RECORDS WERE REVIEWED. ALL PROCESS AND FINAL INSPECTIONS MEET SPECIFICATIONS. AN IN-DEPTH INVESTIGATION AS PART OF THAT CAPA HAS BEEN PERFORMED TO IDENTIFY THE CAUSE OF THE SURGIPHOR BOTTLE CRACKING DURING USE. AT THIS TIME THIS INVESTIGATION DISCOVERED THAT THE COMBINATION OF THE FOLLOWING VARIABLES ARE PROBABLE CAUSES THAT CONTRIBUTE TO THE FAILURE MODE OF SURGIPHOR CONTAINERS CRACKING DURING USE: AGING, BOTTLE WALL THICKNESS/BOTTLE WEIGHT, INSPECTED PRODUCT RETURNED INTO THE MANUFACTURING LOT, AND UPPER END OF THE GAMMA STERILIZATION DOSAGES OF THE PRODUCT. A FOLLOW-UP WILL BE SENT IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 0

A BOTTLE BURST WHILE BEING SQUEEZED DURING APPLICATION OF PRODUCT ON PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1535057 BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM SURGIPHOR FQH CAREFUSION, INC 910100 1854590 00382909101003

Patients

Seq Age Sex Outcome Treatment
1 Unknown