FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 16143548 · Received January 12, 2023

Report

Report Number
9610877-2023-50256
Event Type
Malfunction
Date Received
January 12, 2023
Date of Event
December 19, 2022
Report Date
January 12, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
UDI-DI
04961333234066
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT; THEREFORE, 510K IS NOT APPLICABLE. MODEL: EC34-I10TL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER: K180285. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVER PCB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD DRIVER PCB. IN ADDITION, WE CONFIRMED THAT THE ELECTRICAL CONNECTOR FLUID DAMAGE, THE LG CABLE CONNECTOR FLUID DAMAGE, AND THE ANGLE WIRE PLAY; HOWEVER, THEY ARE NOT THE MAIN CAUSE AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT, ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
634885 PENTAX VIDEO COLONOSCOPE FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC34-I10TL 04961333234066

Patients

Seq Age Sex Outcome Treatment
1 Unknown