PENTAX
Report
- Report Number
- 9610877-2023-50256
- Event Type
- Malfunction
- Date Received
- January 12, 2023
- Date of Event
- December 19, 2022
- Report Date
- January 12, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDF
- UDI-DI
- 04961333234066
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT; THEREFORE, 510K IS NOT APPLICABLE. MODEL: EC34-I10TL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER: K180285. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVER PCB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD DRIVER PCB. IN ADDITION, WE CONFIRMED THAT THE ELECTRICAL CONNECTOR FLUID DAMAGE, THE LG CABLE CONNECTOR FLUID DAMAGE, AND THE ANGLE WIRE PLAY; HOWEVER, THEY ARE NOT THE MAIN CAUSE AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT, ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 634885 | PENTAX | VIDEO COLONOSCOPE | FDF | HOYA CORPORATION PENTAX TOKYO OFFICE | EC34-I10TL | 04961333234066 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |