FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE

MDR report key: 16142090 · Received January 11, 2023

Report

Report Number
1920898-2022-00922
Event Type
Malfunction
Date Received
January 11, 2023
Date of Event
December 22, 2022
Report Date
February 22, 2023
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K212499
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (1) 3/10CC, 8MM, 31G SYRINGE IN AN OPEN POLY BAG FROM LOT # 0203534. CUSTOMER STATES THAT THE NEEDLE HUB SEPARATED. THE RETURNED SYRINGE WAS EXAMINED AND EXHIBITED THE HUB ASSEMBLY SEPARATED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0203534 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE SYRINGE AND REMAINED IN THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CONSUMER RETURNED MY CALL, STATED THAT THE NEEDLE HUB OF ONE SYRINGE SEPARATED AND STAYED INSIDE OF THE SHIELD."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE SYRINGE AND REMAINED IN THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CONSUMER RETURNED MY CALL, STATED THAT THE NEEDLE HUB OF ONE SYRINGE SEPARATED AND STAYED INSIDE OF THE SHIELD."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1418981 BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 0203534

Patients

Seq Age Sex Outcome Treatment
1 Unknown