FDA Adverse Event Injury Summary report: N

IMP,TSV,6.0,10,MTX,MG

MDR report key: 16132741 · Received January 10, 2023

Report

Report Number
0002023141-2023-00098
Event Type
Injury
Date Received
January 10, 2023
Date of Event
October 1, 2021
Report Date
April 21, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K101977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-00996, 0097, 0100. ZIMMER BIOMET COMPLAINT NUMBER: (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. BRAND NAME UNKNOWN / NOT PROVIDED. CATALOG AND LOT NUMBER UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER NOT AVAILABLE. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2023-00096-1, 00097-1 AND 00100-1. CORRECTION: THE INITIAL ASSOCIATED REPORTS: 0002023141-2022-00096, 00097 AND 00100 WERE DOCUMENTED INCORRECTLY WITH THE WRONG YEAR (2022). THE ASSOCIATED REPORTS SHOULD STATE, 0002023141-2023-00096, 00097 AND 00100. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. DHR REVIEW AND COMPLAINT HISTORY REVIEW COULD NOT BE PERFORMED, AS THE LOT NUMBER ASSOCIATED WITH THE REPORTED PRODUCT IS NOT AVAILABLE. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT IS WITHIN SPECIFICATIONS. A COMPLAINT HISTORY REVIEW BY ITEM NUMBER WAS CONDUCTED FOR THE TSVTWB10, TSVT6B8, TSVTB8, AND TSVT6B10 DATING BACK TO 12 MONTHS FROM THE COMPLAINT NOTIFICATION DATE. THE COMPLAINT HISTORY REVIEW REVEALED THAT THERE ARE NO EXISTING NON-CONFORMANCES/CAPA/HHE/D/IE/PRODUCT HOLDS FOR THE REPORTED DEVICE RELATED TO THE REPORTED EVENT. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO MEDICAL CONDITIONS / PATIENT HABITS AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION AND NO ASSOCIATION BETWEEN THE DENTAL IMPLANT AND BONE LOSS WAS FOUND THAT CAN EXPLAIN THESE EVENTS. THEREFORE, THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANTS ON (4) UNKNOWN TEETH SITES WERE REMOVED DUE TO BONE LOSS.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204402 IMP,TSV,6.0,10,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL

Patients

Seq Age Sex Outcome Treatment
1 38 YR Female Required Intervention