IMMULITE 2000
Report
- Report Number
- 2247117-2010-00004
- Event Type
- Other
- Date Received
- February 11, 2010
- Date of Event
- January 13, 2010
- Report Date
- January 13, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- DGC
- PMA / PMN Number
- K970227
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
A SIEMENS HEALTHCARE FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE FOR INSTRUMENT EVAL. ANALYSIS OF THE INSTRUMENT AND THE INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE DISCORDANT HCG RESULTS CANNOT BE DETERMINED. AFTER EVAL OF THE INSTRUMENT, THE FSE REPLACED BOTH PROBE ASSEMBLIES, AND DECONTAMINATED THE FLUIDIC AND SUBSTRATE SYSTEMS. THE INSTRUMENT IS PERFORMING WITHIN SPECS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
DISCORDANT LOW HCG RESULTS WERE OBTAINED ON 3 PATIENT SAMPLES. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED ON THE LAB'S SECOND IMMULITE 2000 SYSTEM (B)(4). THE REPEAT RESULTS FROM (B)(4) WERE REPORTED TO THE PHYSICIAN. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WERE NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HCG RESULTS.
A SIEMENS HEALTHCARE FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE FOR INSTRUMENT EVAL. ANALYSIS OF THE INSTRUMENT AND THE INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE DISCORDANT HCG RESULTS CANNOT BE DETERMINED. AFTER EVAL OF THE INSTRUMENT, THE FSE REPLACED BOTH PROBE ASSEMBLIES, AND DECONTAMINATED THE FLUIDIC AND SUBSTRATE SYSTEMS. THE INSTRUMENT IS PERFORMING WITHIN SPECS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
A SIEMENS HEALTHCARE FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE FOR INSTRUMENT EVAL. ANALYSIS OF THE INSTRUMENT AND THE INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE DISCORDANT HCG RESULTS CANNOT BE DETERMINED. AFTER EVAL OF THE INSTRUMENT, THE FSE REPLACED BOTH PROBE ASSEMBLIES, AND DECONTAMINATED THE FLUIDIC AND SUBSTRATE SYSTEMS. THE INSTRUMENT IS PERFORMING WITHIN SPECS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
DISCORDANT LOW HCG RESULTS WERE OBTAINED ON 3 PATIENT SAMPLES. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED ON THE LAB'S SECOND IMMULITE 2000 SYSTEM (B)(4). THE REPEAT RESULTS FROM (B)(4) WERE REPORTED TO THE PHYSICIAN. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WERE NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HCG RESULTS.
DISCORDANT LOW HCG RESULTS WERE OBTAINED ON 3 PATIENT SAMPLES. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED ON THE LAB'S SECOND IMMULITE 2000 SYSTEM (B)(4). THE REPEAT RESULTS FROM (B)(4) WERE REPORTED TO THE PHYSICIAN. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WERE NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HCG RESULTS.
A SIEMENS HEALTHCARE FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE FOR INSTRUMENT EVAL. ANALYSIS OF THE INSTRUMENT AND THE INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE DISCORDANT HCG RESULTS CANNOT BE DETERMINED. AFTER EVAL OF THE INSTRUMENT, THE FSE REPLACED BOTH PROBE ASSEMBLIES, AND DECONTAMINATED THE FLUIDIC AND SUBSTRATE SYSTEMS. THE INSTRUMENT IS PERFORMING WITHIN SPECS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
A SIEMENS HEALTHCARE FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE FOR INSTRUMENT EVAL. ANALYSIS OF THE INSTRUMENT AND THE INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE DISCORDANT HCG RESULTS CANNOT BE DETERMINED. AFTER EVAL OF THE INSTRUMENT, THE FSE REPLACED BOTH PROBE ASSEMBLIES, AND DECONTAMINATED THE FLUIDIC AND SUBSTRATE SYSTEMS. THE INSTRUMENT IS PERFORMING WITHIN SPECS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
DISCORDANT LOW HCG RESULTS WERE OBTAINED ON 3 PATIENT SAMPLES. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED ON THE LAB'S SECOND IMMULITE 2000 SYSTEM (B)(4). THE REPEAT RESULTS FROM (B)(4) WERE REPORTED TO THE PHYSICIAN. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WERE NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HCG RESULTS.
DISCORDANT LOW HCG RESULTS WERE OBTAINED ON 3 PATIENT SAMPLES. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED ON THE LAB'S SECOND IMMULITE 2000 SYSTEM (B)(4). THE REPEAT RESULTS FROM (B)(4) WERE REPORTED TO THE PHYSICIAN. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WERE NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HCG RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMULITE 2000 | IMMUNO-ASSAY | DGC | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ||||
| 2 | ||||
| 3 |