FDA Adverse Event Injury Summary report: N

SICAT CLASSICGUIDE

MDR report key: 16130897 · Received January 10, 2023

Report

Report Number
3006098230-2022-00001
Event Type
Injury
Date Received
January 10, 2023
Date of Event
December 16, 2022
Report Date
January 10, 2023
Manufacturer
SICAT GMBH & CO. KG
Product Code
NDP
PMA / PMN Number
NA#S-P#NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING SCENARIO APPEARS TO BE THE MOST PROBABLE: THE PLANNED IMPLANT POSITION #11(23) DOES NOT SEEM TO BE COMPLETELY SURROUNDED BY BONE AT THE PALATAL AND SEEMS TO BE PLACED IN AN AREA OF OBLIQUE BONE STRUCTURE. THE DRILL PROBABLY SLID AT THE BONE STRUCTURE AND CHANGED THE DRILLING DIRECTION. THIS OCCURED BECAUSE THE DENTIST DID NOT PRESS THE SURGICAL GUIDE SUFFICIENTLY TIGHT ONTO THE TEETH DURING SURGERY.

Description of Event or Problem · 0

THE REPORTING DENTIST HAS USED A SICAT SURGICAL GUIDE (SICAT CLASSICGUIDE) FOR PREPARING OSTEOTOMIES (DRILL HOLES FOR ACCOMMODATING A DENTAL IMPLANT) FOR DENTAL IMPLANTS. HOWEVER, THE IMPLANT (TOOTH #11) DID NOT END UP AT THE CORRECT POSITION AND ORIENTATION. THE DENTIST REMOVED THE IMPLANT AND GRAFTED THE SITE AND WILL LET IT HEAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1723575 SICAT CLASSICGUIDE SURGICAL GUIDE NDP SICAT GMBH & CO. KG 10301

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male Required Intervention DENT. IMPL. SWEDEN&MARTINA OUT-LINK2 E2-ZT-330-115.