FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES

MDR report key: 16130477 · Received January 10, 2023

Report

Report Number
1917413-2022-00832
Event Type
Malfunction
Date Received
January 10, 2023
Date of Event
September 15, 2022
Report Date
April 24, 2023
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903679626
PMA / PMN Number
K945952
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO ERRONEOUS RESULTS WERE OBSERVED. 90 RETENTION SAMPLES WERE VISUALLY INSPECTED WITH NO ISSUES BEING IDENTIFIED. ADDITIONALLY, CLINICAL TESTING WAS PERFORMED ON RETENTION SAMPLES: NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES EXHIBITED PROPER FILL. BD WAS UNABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (ERRONEOUS RESULTS-HIGH POTASSIUM) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE RETENTION LOT SAMPLES. REPLICATES OF BOTH RETAIN AND CONTROL SAMPLES TESTED WERE ACCEPTABLE IN TERMS OF BOTH PRECISION AND ACCURACY. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE ERRONEOUS RESULTS. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. FACTORS THAT MAY CONTRIBUTE TO ERRONEOUS RESULTS WERE EVALUATED THROUGH A CLINICAL STUDY TO VERIFY THE DESIGN OF THE DEVICE MET IT¿S INTENDED USE. THE RESULT OF THE STUDY SHOWED THAT THE DEVICE PERFORMED AS EXPECTED AND WE WERE UNABLE TO DETERMINE ANY EXTERNAL CONTRIBUTOR TO THIS REPORTED ISSUE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE THERE WAS ERRONEOUS RESULTS THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WERE "ERRONEOUS RESULTS WITH PRODUCT 367962, 2080690. RESULTS WERE 6.5 AND SENT TO REF LAB TO CONFIRM RESULTS. PATIENT WAS REDRAWN ON 11/21 AND 4.6 RESULTS WERE STATED TO BE NORMAL.".

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE THERE WAS ERRONEOUS RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WERE "ERRONEOUS RESULTS WITH PRODUCT 367962, 2080690. RESULTS WERE 6.5 AND SENT TO REF LAB TO CONFIRM RESULTS. PATIENT WAS REDRAWN ON 11/21 AND 4.6 RESULTS WERE STATED TO BE NORMAL."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255984 BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367962 2080690 50382903679626

Patients

Seq Age Sex Outcome Treatment
1 Unknown