BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES
Report
- Report Number
- 1917413-2022-00832
- Event Type
- Malfunction
- Date Received
- January 10, 2023
- Date of Event
- September 15, 2022
- Report Date
- April 24, 2023
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 50382903679626
- PMA / PMN Number
- K945952
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO ERRONEOUS RESULTS WERE OBSERVED. 90 RETENTION SAMPLES WERE VISUALLY INSPECTED WITH NO ISSUES BEING IDENTIFIED. ADDITIONALLY, CLINICAL TESTING WAS PERFORMED ON RETENTION SAMPLES: NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES EXHIBITED PROPER FILL. BD WAS UNABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (ERRONEOUS RESULTS-HIGH POTASSIUM) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE RETENTION LOT SAMPLES. REPLICATES OF BOTH RETAIN AND CONTROL SAMPLES TESTED WERE ACCEPTABLE IN TERMS OF BOTH PRECISION AND ACCURACY. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE ERRONEOUS RESULTS. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. FACTORS THAT MAY CONTRIBUTE TO ERRONEOUS RESULTS WERE EVALUATED THROUGH A CLINICAL STUDY TO VERIFY THE DESIGN OF THE DEVICE MET IT¿S INTENDED USE. THE RESULT OF THE STUDY SHOWED THAT THE DEVICE PERFORMED AS EXPECTED AND WE WERE UNABLE TO DETERMINE ANY EXTERNAL CONTRIBUTOR TO THIS REPORTED ISSUE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE THERE WAS ERRONEOUS RESULTS THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WERE "ERRONEOUS RESULTS WITH PRODUCT 367962, 2080690. RESULTS WERE 6.5 AND SENT TO REF LAB TO CONFIRM RESULTS. PATIENT WAS REDRAWN ON 11/21 AND 4.6 RESULTS WERE STATED TO BE NORMAL.".
IT WAS REPORTED WHEN USING THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE THERE WAS ERRONEOUS RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE WERE "ERRONEOUS RESULTS WITH PRODUCT 367962, 2080690. RESULTS WERE 6.5 AND SENT TO REF LAB TO CONFIRM RESULTS. PATIENT WAS REDRAWN ON 11/21 AND 4.6 RESULTS WERE STATED TO BE NORMAL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255984 | BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 367962 | 2080690 | 50382903679626 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |