FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1612229 · Received February 4, 2010

Report

Report Number
2027969-2010-00142
Event Type
Malfunction
Date Received
February 4, 2010
Date of Event
January 11, 2010
Report Date
February 4, 2010
Manufacturer
BIOSITE INCORPORATED
Product Code
JPA
PMA / PMN Number
K021923
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B) (6) 2010, INRATIO: 1.0, REFERENCE: 4.0, MEAN: 2.50, CONFIDENCE LIMITS: 1.6-3.4. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES FALL OUTSIDE THE LIMITS, SO THE CRITERIA ARE NOT MET AND THE VALUES ARE DISCREPANT BEYOND THE DOCUMENTED VARIABILITY FOR PT/INR TESTING. STRIP ACCURACY TESTING REQUIRED AS PART OF THE COMPLAINT INVESTIGATION. RETAIN LOT#222166 WITH CODE UA293 HAS BEEN INVESTIGATED. INVESTIGATION RESULTS: ACCURACY, THE ALLOWABLE DIFFERENCE BETWEEN THE INRATIO INR'S AND SYSMEX INR'S ARE CALCULATED. AT LEAST TWO OUT OF THREE REPLICATES FOR BOTH SAMPLES FOR STRIP LOT 222166 ARE WITHIN THE ALLOWABLE BIAS. NO DISCREPANT RESULTS ARE PRODUCED ON IN-HOUSE METERS. THESE MEET THE ABOVE CRITERIA, AND THEN NO FURTHER ACTION IS REQUIRED. CUSTOMER RESULTS ANALYZED AND ACCURACY CONFIDENCE LIMITS WERE NOT MET. TIME ELAPSE BETWEEN RESULTS NOT INDICATED. IF THE TIME ELAPSED EXCEEDS 3 HOURS, THERE IS A HIGH POSSIBILITY THAT THE DISCREPANCIES ARE DUE TO CHANGES IN THE STATUS OF THE PATIENT. IN-HOUSE ACCURACY TEST PERFORMED WITH RETAIN STRIP, IN-HOUSE METER, AND ACCURACY CRITERIA WERE MET. PRODUCT DEFICIENCY NOT ESTABLISHED. NO FURTHER INVESTIGATION WILL BE PURSUED. AS OF 2/04/2010, 1 DISCREPANT RESULT COMPLAINT WAS REPORTED FOR LOT # 222166 YIELDING A COMPLAINT RATE OF 0.005%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. ON GOING TRENDING SHALL BE PERFORMED TO DETERMINE IF FURTHER ACTION IS WARRANTED. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B) (6) 2010, INRATIO: APPROX 1.0; LAB: APPROX 4.0. CALLER DID NOT KNOW EXACT INR RESULTS, PATIENT SPECIFICS, OR TIME INTERVAL BETWEEN METER TEST AND LAB DRAW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA BIOSITE INCORPORATED 100139 222166

Patients

Seq Age Sex Outcome Treatment
1