FDA Adverse Event
Malfunction
Summary report: N
OIL FOR EMBRYO CULTURE
MDR report key: 16120606
·
Received January 9, 2023
Report
- Report Number
- 16120606
- Event Type
- Malfunction
- Date Received
- January 9, 2023
- Date of Event
- December 6, 2022
- Report Date
- December 21, 2022
- Manufacturer
- FUJIFILM IRVINE SCIENTIFIC, INC.
- Product Code
- MQL
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MA
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
Description of Event or Problem · 0
EMBRYOLOGIST NOTICED DEGENERATION OF EMBRYOS TREATED WITH MINERAL OIL USED TO PREVENT EVAPORATION. THE EMBRYOLOGIST THEN PERFORMED A QA/QI TEST OF THE MINERAL OIL ON MOUSE EMBRYOS AND NOTED SAME EFFECT. APPROXIMATELY 20 PATIENTS MAY BE AFFECTED. THE OIL HAS BEEN REMOVED FROM USE. MANUFACTURER RESPONSE FOR MINERAL OIL, OIL FOR EMBRYO CULTURE (PER SITE REPORTER) NO RESPONSE TO DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1394814 | OIL FOR EMBRYO CULTURE | MEDIA, REPRODUCTIVE | MQL | FUJIFILM IRVINE SCIENTIFIC, INC. | 9305-100ML | 0000011351 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female |