- Report Number
- 2032227-2023-103238
- Event Type
- Malfunction
- Date Received
- January 9, 2023
- Date of Event
- January 1, 2023
- Report Date
- March 16, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000187422
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
RETAINER RING = CLEAR. CASE TYPE = NGP CUSTOMER RETURNED PUMP FOR AN ALLEGED CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM FOUND ON (B)(6) 2023. NO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM NOTED DURING BOOT UP. THE PUMP PASSED THE DISPLACEMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, AND ACTIVE CURRENT TEST. PUMP FAILED SELF TEST DUE TO VIBRATE ANOMALY. PUMP FAILS SLEEP CURRENT TEST. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS AND CARELINK. THE FOLLOWING ALARMS WERE NOTED ON EVENT DATE: PUMP ERROR 4 ON (B)(6) 2023 14:26:01.000 AND PUMP ERROR 63 ON (B)(6) 2023 14:26:02.000. CONFIRMED PUMP ERROR 4 (LINENUMBER= 2727 FILENUMBER= 32122) DUE TO HARDWARE ERROR. CONFIRMED PUMP ERROR 63 VARIABLE 21 DUE TO HARDWARE ERROR. HARDWARE ERROR ISOLATED TO ELECTRONIC STACK. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON PCBA 1, MOTOR, FORCE SENSOR, KEYPAD CONNECTOR, AND LITHIUM BATTERY CONNECTOR. NO FATAL CRITICAL ALARMS OR THREE CONSECUTIVE CRITICAL HANDLING ALARMS FOUND IN THE HISTORY FILES OR TRACES. UNABLE TO PERFORM FORCE SENSOR ZERO OFFSET DUE TO MOISTURE DAMAGE ON FORCE SENSOR CONNECTOR. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: BATTERY TUBE THREADS - CRACKED, STAINED KEYPAD OVERLAY, CRACKED CASE-CORNER OF BELT CLIP RAILS, PILLOWING KEYPAD OVERLAY, KEYPAD OVERLAY TEXTURE DAMAGE, PARTIALLY FADED SERIAL NUMBER LABEL, CRACKED RETAINER, PARTIALLY DETACHED RETAINER. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS NOT OBSERVED DURING TESTING. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM NOT CONFIRMED. PUMP ERROR 4 AND PUMP ERROR 63 CONFIRMED DUE TO HARDWARE ERROR, ISOLATED TO ELECTRONIC STACK. VIBRATE ANOMALY CONFIRMED DUE TO MOISTURE DAMAGE. SLEEP CURRENT ANOMALY CONFIRMED DUE TO MOISTURE DAMAGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED CRITICAL PUMP ERROR. CUSTOMER REPORTED THAT THEY RECEIVED THE OPEN BOOK IMAGE ON THE INSULIN PUMP SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS SUCCESSFULLY PERFORMED; HOWEVER, THE CUSTOMER WILL DISCONTINUE USE OF THE DEVICE.