FDA Adverse Event Other Summary report: N

TOTALCARE BARIATRIC

MDR report key: 1610963 · Received February 10, 2010

Report

Report Number
1824206-2010-02085
Event Type
Other
Date Received
February 10, 2010
Date of Event
January 12, 2010
Report Date
January 12, 2010
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND, THE ACCOUNT ROUTED THE PENDANT CABLE UNDER THE BED FRAME. WHEN THE BED WAS LOWERED THE PENDANT CABLE WAS CUT WHICH EXPOSED BARE WIRING. THE ACCOUNT DECLINED TO RELEASE MORE INFO ABOUT THE STAFF MEMBER INVOLVED IN THE EVENT.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THAT THE BED WAS IN THE OPERATING ROOM WHEN A PT WAS PLACED ON THE BED FROM SURGERY. THE HAND CONTROL WAS STUCK IN THE SIDERAIL AND WHEN THE NURSE PULLED THE HAND CONTROL, SHE TOUCHED AN OPEN WIRE ON THE CORD. THE ACCOUNT ALLEGED, THE NURSE WAS BURNED BUT DECLINED TO RELEASE INFO REGARDING THE NURSE INVOLVED IN THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BARIATRIC AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1840

Patients

Seq Age Sex Outcome Treatment
1 UNK