FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 3 ML LL EURO 200 S/C

MDR report key: 16109210 · Received January 5, 2023

Report

Report Number
1213809-2022-01342
Event Type
Malfunction
Date Received
January 5, 2023
Date of Event
November 23, 2022
Report Date
July 17, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096580
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT WAS OBSERVED THAT THE LIQUID PRESENT PAST THE STOPPER AND ACCUMULATED AROUND THE PLUNGER ROD. POTENTIAL ROOT CAUSE FOR THE LEAKAGE PAST STOPPER DEFECT COULD NOT ESTABLISHED AT THIS TIME. THE PHYSICAL SAMPLE IS REQUIRED FOR MORE IN DEPTH INVESTIGATION OF ALL POTENTIAL CONTRIBUTING FACTORS. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2221955. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD SYRINGE 3 ML LL EURO 200 S/C LEAKAGE AND UNDERDOSAGE OCCURRED TWICE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE MEDICATION LIQUID GOT BEHIND THE PISTON. PATIENT WOULD HAVE RECEIVED TOO LITTLE MEDICATION DOSAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1313637 BD SYRINGE 3 ML LL EURO 200 S/C PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309658 2221955 30382903096580

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown