PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2023-00155
- Event Type
- Injury
- Date Received
- January 5, 2023
- Date of Event
- December 15, 2022
- Report Date
- March 8, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VM
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
THE DEVICE WAS RETURNED FOR ANALYSIS. THE SUTURE MATERIAL SEPARATION WAS CONFIRMED AS A LINK SEPARATION. A REVIEW OF THE MANUFACTURING RECORDS IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE CAUSE OF THE REPORTED DIFFICULTY AND THE SUBSEQUENT TREATMENT ARE DUE TO THE CIRCUMSTANCES OF THE PROCEDURE. IN THIS CASE, IT IS LIKELY, THE SEPARATION IS RELATED TO AN INTERACTION OF THE DEVICE WITH THE SUTURE PULLED UNTIL IT BREAKS, OR TENSIONING ANGLE IS NOT BEING COAXIAL. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D4: LOT NUMBER CHANGED FROM 2101141 TO 0112541.
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY USING A PROGLIDE AFTER A CAROTID INTERVENTIONAL PROCEDURE USING A 6F SHEATH. REPORTEDLY, WHEN REMOVING THE PLUNGER IT WAS NOTED THAT THE SUTURE BROKE. ANOTHER PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.