BONE HOOK-SHAPE ARM 206MM
Report
- Report Number
- 8030965-2023-00134
- Event Type
- Malfunction
- Date Received
- January 4, 2023
- Date of Event
- December 9, 2022
- Manufacturer
- SYNTHES GMBH
- Product Code
- LXH
- UDI-DI
- 10886982201591
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: H3, H4, H6: THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THERE WAS NO DAMAGE OR DEFECTS WITH THE BONE HOOK-SHAPE ARM 206MM, P/N: 398.756, ONLY SIGNS OF NORMAL USAGE. A DIMENSIONAL INSPECTION WAS UNABLE TO BE PERFORMED DUE TO IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A FUNCTIONAL TEST WAS UNABLE TO PERFORMED DUE TO MATING DEVICE WAS NOT RETURNED, IN ADDITION BUTTON SHAFT WAS PRESSED AND THE INTERLOCKING MECHANISM WORKS AS EXPECTED. THE COMPLAINT CONDITION NOT ABLE TO BE REPLICATED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE BONE HOOK-SHAPE ARM 206MM, P/N: 398.756 WAS FOUND TO HAVE NO DAMAGE OR DEFECTS. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DRAWING/SPECIFICATIONS REVIEWED? YES, REVIEWED. DEVICE HISTORY LOT: PART # 398.756, SYNTHES LOT # 09-1506, SUPPLIER LOT # 09-1506, RELEASE TO WAREHOUSE DATE # 05 MAR 2010, SUPPLIER # SYNTHES GMBH OBERDORF. NO NCRS WERE GENERATED DURING PRODUCTION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DATE OF CONCOMITANT THERAPY IS (B)(6) 2022.
PRODUCT COMPLAINT#: (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: UN-RECOGNIZED LOT NUMBER: 190438/401. LOT#: PROVIDED IS NOT A VALID NUMBER, THEREFORE, THE DHR COULD NOT BE COMPLETED. IF DEVICE IS RETURNED OR LOT NUMBER CAN BE CONFIRMED, THE DHR WILL BE REVISITED. DEVICE AVAILABLE FOR EVALUATION: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D4: EXPIRATION DATE ADDED H4: DEVICE MANUFACTURE DATE ADDED H11: CORRECTED DATA- D9 DEVICE RETURNED TO MANUFACTURER. D4: LOT NUMBER CORRECTED. H3, H6: THE DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. D9, D4.
IT WAS REPORTED THAT ON (B)(6) 2022, THE HOOKS DID NOT STAY COMPRESSED WHEN CLAMPED. THE SURGICAL DELAY, ADDITIONAL INTERVENTION AND PATIENT CONSEQUENCE WAS UNKNOWN. THE SURGERY WAS SUCCESSFULLY COMPLETED. THIS COMPLAINT INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) BONE HOOK-SHAPE ARM 206MM. THIS IS REPORT 2 OF 3 FOR COMPLAINT: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1191765 | BONE HOOK-SHAPE ARM 206MM | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | SYNTHES GMBH | 398.756 | 09-1506 | 10886982201591 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | COLLINEAR REDUCTION CLAMP SLIDING MECHAN| PERCUTANEOUS ARM 255MM |