BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Report
- Report Number
- 1917413-2022-00818
- Event Type
- Malfunction
- Date Received
- January 4, 2023
- Date of Event
- December 7, 2022
- Report Date
- May 4, 2023
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 00382903627608
- PMA / PMN Number
- K911731
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 2 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS NOT OBSERVED. NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES EXHIBITED PROPER FILL. BD WAS UNABLE TO DUPLICATE THE CUSTOMER¿S INDICATED FAILURE (POOR BARRIER) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. REPLICATES OF BOTH RETENTION AND CONTROL SAMPLES TESTED DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE FOR ALL VISUAL OBSERVATIONS EVALUATED. ALL TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER SEPARATION BASED ON PHOTOS ONLY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, THE DEVICE EXPERIENCED POOR BARRIER SEPARATION OF THE SAMPLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: CUSTOMER REPOTS RECURRING SEPARATION ISSUES WHEN USING CAT 362760 LOT 2173885.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, THE DEVICE EXPERIENCED POOR BARRIER SEPARATION OF THE SAMPLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: CUSTOMER REPOTS RECURRING SEPARATION ISSUES WHEN USING CAT 362760 LOT 2173885.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1118637 | BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 2173885 | 00382903627608 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |