FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE

MDR report key: 16093196 · Received January 4, 2023

Report

Report Number
9610773-2023-00067
Event Type
Malfunction
Date Received
January 4, 2023
Date of Event
December 1, 2021
Report Date
January 4, 2023
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K111788
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO OLYMPUS HONGKONG & CHINA LTD. (OHC), (RECEIVED AT OHC ON 2022-12-07). THE INSPECTION AT OHC DETECTED VARYING COLOURED IMAGES (PURPLE/GREEN) AND TRACED THE FAILURE BACK TO A DEFECTIVE R-UNIT. THEREFORE, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. THE REPORTED LEAKINESS COULD NOT BE CONFIRMED. THE AREA UNDER THE HANDLE (AS WELL AS UNDER THE INNER BODY) IS NOT A SEALED VOLUME. CONSEQUENTLY, MOISTURE IN THE REPORTED AREAS DO NOT INDICATE A DEFICIENCY OF THE DEVICE. A MATERIAL OR QUALITY PROBLEM CAN BE EXCLUDED SINCE A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE VIDEO TELESCOPE WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED ON OLYMPUS SIDE WITH NO FURTHER ACTIONS. THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES AND THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS.

Description of Event or Problem · 0

THE LOANER VIDEO TELESCOPE REPORTEDLY DISPLAYED A PURPLE IMAGE AND SHOWED LEAKINESS AT THE ASSEMBLY OF THE OUTER TUBE AND R-UNIT. NO PATIENT/PROCEDURE WAS INVOLVED.