FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 16091169 · Received January 3, 2023

Report

Report Number
3002808148-2023-00040
Event Type
Malfunction
Date Received
January 3, 2023
Date of Event
December 5, 2022
Report Date
February 2, 2023
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
UDI-DI
04953170363498
PMA / PMN Number
K102379
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE ROOT CAUSE OF THE NO IMAGE COULD NOT BE DETERMINED AS THE ISSUE WAS NOT REPRODUCED. POSSIBLE CAUSES FOR THE ¿NO IMAGE¿ MIGHT HAVE BEEN DUE TO A FAULTY CIRCUIT BOARD (QB), A HARNESS CONNECTION FAILURE WITH THE LIQUID CRYSTAL DISPLAY (LCD) PANEL, OR A FAULTY LCD PANEL. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE SUBJECT DEVICE HAS BEEN RETURNED TO OLYMPUS FOR EVALUATION AND THE REPORTED ISSUE WAS NOT CONFIRMED. DURING EVALUATION, IT WAS OBSERVED, THE SCREEN IF THE SUBJECT DEVICE WAS SCRATCHED. THE INVESTIGATION IS ONGOING, AND A DEFINITIVE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS, DURING REPROCESSING FOR A DIAGNOSTIC GASTROSCOPIC PROCEDURE, THE HIGH-DEFINITION LCD MONITOR HAD NO IMAGE. SUBSEQUENTLY, THE INTENDED PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146183 HIGH DEFINITION LCD MONITOR HIGH DEFINITION LCD MONITOR FET SHIRAKAWA OLYMPUS CO., LTD. OEV262H 04953170363498

Patients

Seq Age Sex Outcome Treatment
1 Unknown MODEL NUMBER: CV-290, NO SERIAL NUMBER PROVIDED