FDA Adverse Event Malfunction Summary report: N

SKATER DRAINAGE CATHETER - LOCKING PIGTAIL 8F X 25CM

MDR report key: 16089825 · Received January 3, 2023

Report

Report Number
0001625425-2023-00918
Event Type
Malfunction
Date Received
January 3, 2023
Date of Event
December 4, 2022
Report Date
February 13, 2023
Manufacturer
ARGON MEDICAL DEVICES
Product Code
OAJ
UDI-DI
00886333000159
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE SAMPLE HAS BEEN RECEIVED AND REVIEWED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE MANUFACTURING AND INSPECTION RECORDS FOR THIS LOT WAS CONDUCTED, AND NO DEVIATIONS AND NON-CONFORMANCES WERE RECORDED. A REVIEW OF RETURNED PRODUCT FROM THE CUSTOMER WAS PERFORMED AND VISUALLY INSPECTED. THE SAMPLE RETURNED FROM THE CUSTOMER HAD THE LOCKING STRING BROKEN AND THE COMPLAINT WAS CONFIRMED. PACKAGE WAS OPEN AND SNAP OFF TAB AND STRING WAS LOOSE INSIDE THE PACKAGE. THE MOST PROBABLE ROOT CAUSE COULD BE THAT USER USED EXCESSIVE FORCE OR IMPROPER ANCHOR DURING THE PROCEDURE. THE TRO-CAR NEEDLE COULD HAVE ADVANCED TO THE ON POSITION AND CUT/DAMAGED THE STRING. SINCE THE BREAKAGE COULD BE DUE TO AN EVENT WITHIN THE USER ENVIRONMENT. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 0

THE PATIENT'S LIVER ABSCESS WAS PUNCTURED AND DRAINED, AND THE CATHETER WAS EXTUBATED NOW. THE DRAINAGE TUBE WAS PULLED OUT SMOOTHLY, BUT THE LEAD WAS NOT BROUGHT OUT SMOOTHLY. THE SURGEON WAS ASKED TO ASSIST IN DIAGNOSIS AND TREATMENT. THE RESISTANCE WAS GREAT WHEN TRYING TO PULL THE LEAD FOR MANY TIMES. FINALLY, THE LEAD WAS PARTIALLY BROKEN AND SOME REMAINED IN THE PATIENT'S BODY.

Description of Event or Problem · 0

THE PATIENT'S LIVER ABSCESS WAS PUNCTURED AND DRAINED, AND THE CATHETER WAS EXTUBATED NOW. THE DRAINAGE TUBE WAS PULLED OUT SMOOTHLY, BUT THE LEAD WAS NOT BROUGHT OUT SMOOTHLY. THE SURGEON WAS ASKED TO ASSIST IN DIAGNOSIS AND TREATMENT. THE RESISTANCE WAS GREAT WHEN TRYING TO PULL THE LEAD FOR MANY TIMES. FINALLY, THE LEAD WAS PARTIALLY BROKEN AND SOME REMAINED IN THE PATIENT'S BODY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146272 SKATER DRAINAGE CATHETER - LOCKING PIGTAIL 8F X 25CM SKATER OAJ ARGON MEDICAL DEVICES 755108025 11421909 00886333000159
1606476 SKATER DRAINAGE CATHETER - LOCKING PIGTAIL 8F X 25CM SKATER OAJ ARGON MEDICAL DEVICES 755108025 11421909 00886333000159

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other| R