FDA Adverse Event
Malfunction
Summary report: N
CARDIOSAVE HYBRID, TYPE B PLUG
MDR report key: 16074887
·
Received December 29, 2022
Report
- Report Number
- 2249723-2022-03509
- Event Type
- Malfunction
- Date Received
- December 29, 2022
- Report Date
- December 21, 2022
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
A GETINGE SERVICE TERRITORY MANAGER (STM) WAS DISPATCHED TO EVALUATE THE IABP UNIT. THE STM FOUND INTRUSION INTO THE SAFETY DISK AND DRIPPING ON THE PNEUMATIC INTERFACE PCB. THE PNEUMATIC MODULE ASSEMBLY (0997-00-1178) AND PNEUMATIC INTERFACE PCB (0670-00-1167) WERE REPLACED. THE STM COMPLETED ALL SAFETY, FUNCTIONALITY AND CALIBRATION CHECKS AND ALL TESTS PASSED TO FACTORY SPECIFICATIONS. THE IABP UNIT WAS CLEARED FOR CLINICAL USE AND RELEASED TO THE CUSTOMER. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA (B)(4).
Description of Event or Problem · 0
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) EXPERIENCED BLOOD BACK INTO PUMP. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2410549 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Male |