FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID, TYPE B PLUG

MDR report key: 16074887 · Received December 29, 2022

Report

Report Number
2249723-2022-03509
Event Type
Malfunction
Date Received
December 29, 2022
Report Date
December 21, 2022
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567108391
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A GETINGE SERVICE TERRITORY MANAGER (STM) WAS DISPATCHED TO EVALUATE THE IABP UNIT. THE STM FOUND INTRUSION INTO THE SAFETY DISK AND DRIPPING ON THE PNEUMATIC INTERFACE PCB. THE PNEUMATIC MODULE ASSEMBLY (0997-00-1178) AND PNEUMATIC INTERFACE PCB (0670-00-1167) WERE REPLACED. THE STM COMPLETED ALL SAFETY, FUNCTIONALITY AND CALIBRATION CHECKS AND ALL TESTS PASSED TO FACTORY SPECIFICATIONS. THE IABP UNIT WAS CLEARED FOR CLINICAL USE AND RELEASED TO THE CUSTOMER. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE ON A PATIENT, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) EXPERIENCED BLOOD BACK INTO PUMP. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2410549 CARDIOSAVE HYBRID, TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-53 10607567108391

Patients

Seq Age Sex Outcome Treatment
1 69 YR Male