630G INSULIN PUMP MMT-1715KL 630G
Report
- Report Number
- 2032227-2022-391116
- Event Type
- Injury
- Date Received
- December 29, 2022
- Date of Event
- December 9, 2022
- Report Date
- January 24, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- 003
Narratives
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. ADDITIONAL INFORMATION HAS BEEN RECEIVED WHICH WAS NOT INCLUDED WITH THE INITIAL REPORT. THE INFORMATION HAS BEEN PROVIDED IN SECTION B5 WITH THIS REPORT. RETAINER RING = BLACK CASE TYPE = NGP CUSTOMER RETURNED PUMP FOR AN ALLEGED WAS HOSPITALIZED FOR HIGH BGS, LOW BGS AND DKA FOUND ON (B)(6) 2022. ALSO, ON (B)(4) DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM, BOLUS DELIVERY ANOMALY, POSSIBLE UNDER DELIVERY ANOMALY AND WAS HOSPITALIZED FOR HIGH BGS AND DKA FOUND ON (B)(6) 2022. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08715 INCHES. NO UNEXPECTED OCCLUSIONS OR INSULIN FLOW BLOCKED ALARM NOTED DURING TESTING. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. SUCCESSFULLY UPLOADED PUMP TO CARELINK. THERE WAS ¿NO¿ NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM RECORDED IN THE FORMATTED HISTORY FILE FOR THE EVENT DATE. IN FURTHER FULL REVIEW OF THE PUMP HISTORY/TRACES ON THE EVENT DATE OF 09-DEC-2022, THERE IS NO UNEXPECTED ALARMS/SUSPENDS AND FOUND BOLUS DELIVERY OF DAILYTOTALOFBOLUSINSULINDELIVERED = 17.9 U. 12/09/2022 10:30:22.000 NORMALBOLUSDELIVERED NORMALBOLUSAMOUNTPROGRAMMED = 1.5 BOLUSAMOUNT DELIVERED = 1.5 12/09/2022 14:22:22.000 NORMALBOLUSDELIVERED NORMALBOLUSAMOUNTPROGRAMMED = 7.1 BOLUSAMOUNTDELIVERED = 7.1 12/09/2022 16:20:48.000 NORMALBOLUSDELIVERED NORMALBOLUS AMOUNT PROGRAMMED = 6 BOLUSAMOUNTDELIVERED = 6 12/09/2022 18:10:18.000 NORMALBOLUSDELIVERED NORMAL BOLUS AMOUNTPROGRAMMED = 3.3 BOLUSAMOUNTDELIVERED = 3.3 THE PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL BOLUS DELIVERED PROPERLY THEIR INDICATED AMOUNTS (AT QUICK BOLUS SPEED) AND WERE PROPERLY RECORDED IN THE DAILY HISTORY. NO BOLUS DELIVERY ANOMALY OR HISTORY ANOMALY NOTED. NO UNDER DELIVERY ANOMALY OR OVER DELIVERY ANOMALY NOTED DURING TESTING. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA1, PCBA2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. FORCE SENSOR ZERO OFFSET WITHIN SPECIFICATION (23.5 MV). THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE ON THE NGP STB3 AND PASSED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A SCRATCHED CASE, A CRACKED CASE BEHIND THE PUMP NEAR THE BATTERY TUBE COMPARTMENT AND A SERIAL NUMBER LABEL FADING. THE PUMP PASSED ALL THE REQUIRED TESTING. UNABLE TO VERIFY CUSTOMER ALLEGED FOR LOW BGS, HIGH BGS AND DKA. THE FORCE SENSOR IS WITHIN SPECIFICATION AND THE MOTOR FUNCTIONING PROPERLY. CUSTOMER ALLEGED FOR POSSIBLE UNDER DELIVERY WAS NOT CONFIRMED. NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM AND BOLUS DELIVERY ANOMALY WERE NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. INITIAL REPORT WAS SUBMITTED WITH MISSING INFORMATION. THE CORRECTED INFORMATION HAS BEEN UPDATED AND PROVIDED IN SECTION B5 AND H6.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. MEDICAL DEVICE PROBLEM CODE HAS BEEN UPDATED AND PROVIDE WITH THIS REPORT IN SECTION H6.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
BLOOD GLUCOSE VALUE WAS WENT FROM 400 TO 500 MG/DL. THE DEVICE WAS RETURNED FOR ANALYSIS.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS ON UNKNOWN DATE AND EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 400 MG/DL AT THE TIME OF THE EVENT. CUSTOMER HAD SYMPTOMS SUCH AS HEADACHE AND THIRSTY. TROUBLESHOOTING WAS NOT PERFORMED. IT WAS UNKNOWN WHETHER THE CUSTOMER WAS USING THE INSULIN PUMP WITHIN 48 HOURS OF THE EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND IT WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2825277 | 630G INSULIN PUMP MMT-1715KL 630G | ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715KL | NG3212848H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Female | Hospitalization | FRN-MMT-332-RSVR, UNOMED INF SET |