FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID, TYPE B PLUG

MDR report key: 16065421 · Received December 28, 2022

Report

Report Number
2249723-2022-03467
Event Type
Malfunction
Date Received
December 28, 2022
Date of Event
September 11, 2020
Report Date
December 21, 2022
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567108391
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A GETINGE SERVICE TERRITORY MANAGER (STM) WAS DISPATCHED TO INVESTIGATE. THE STM EVALUATED THE IABP UNIT AND TRACED BLOOD FROM THE PNEUMATIC MODULE ASSEMBLY (PIM) TO THE SCROLL COMPRESSOR. THE STM REMOVED THE CONTAMINATED PARTS AND REPAIRS ARE ON HOLD UNTIL THE REPAIR PURCHASE ORDER IS APPROVED. SUBSEQUENTLY, THE STM RETURNED TO THE CUSTOMER'S SITE AT A LATER DATE AND PERFORMED THE REPAIRS ON THE IABP UNIT. THE STM REPLACED THE PNEUMATIC MODULE ASSEMBLY (0997-00-1178), THE PIM TO BACKPLANE CABLE ASSEMBLY (0012-00-1640), TIDAL VOLUME DISK (0202-00-0142), HELIUM RESERVOIR ASSEMBLY (0997-00-0565), PRESSURE-VACUUM RESERVOIR (0202-00-0170-03), MUFFLER TO RESERVOIR FITTING (0103-00-0626), O-RING (0354-00-0175), RELATED FILTERS (0103-00-0499 & 0103-00-0709), IN-LINE FEMALE PNEUMATIC FITTING (0103-00-0657-01), TUBING (0008-00-0377), AND RELATED SCREW KITS. THE STM THEN PERFORMED PREVENTATIVE MAINTENANCE (PM) INCLUDING ALL SAFETY, FUNCTIONALITY, AND CALIBRATION CHECKS AND ALL TESTS PASSED TO FACTORY SPECIFICATIONS. THE IABP UNIT WAS CLEARED FOR CLINICAL USE AND RELEASED TO THE CUSTOMER. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD BLOOD BACK INTO THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
453526 CARDIOSAVE HYBRID, TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-53 10607567108391

Patients

Seq Age Sex Outcome Treatment
1 Unknown