CARDIOSAVE HYBRID, TYPE B PLUG
Report
- Report Number
- 2249723-2022-03467
- Event Type
- Malfunction
- Date Received
- December 28, 2022
- Date of Event
- September 11, 2020
- Report Date
- December 21, 2022
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
A GETINGE SERVICE TERRITORY MANAGER (STM) WAS DISPATCHED TO INVESTIGATE. THE STM EVALUATED THE IABP UNIT AND TRACED BLOOD FROM THE PNEUMATIC MODULE ASSEMBLY (PIM) TO THE SCROLL COMPRESSOR. THE STM REMOVED THE CONTAMINATED PARTS AND REPAIRS ARE ON HOLD UNTIL THE REPAIR PURCHASE ORDER IS APPROVED. SUBSEQUENTLY, THE STM RETURNED TO THE CUSTOMER'S SITE AT A LATER DATE AND PERFORMED THE REPAIRS ON THE IABP UNIT. THE STM REPLACED THE PNEUMATIC MODULE ASSEMBLY (0997-00-1178), THE PIM TO BACKPLANE CABLE ASSEMBLY (0012-00-1640), TIDAL VOLUME DISK (0202-00-0142), HELIUM RESERVOIR ASSEMBLY (0997-00-0565), PRESSURE-VACUUM RESERVOIR (0202-00-0170-03), MUFFLER TO RESERVOIR FITTING (0103-00-0626), O-RING (0354-00-0175), RELATED FILTERS (0103-00-0499 & 0103-00-0709), IN-LINE FEMALE PNEUMATIC FITTING (0103-00-0657-01), TUBING (0008-00-0377), AND RELATED SCREW KITS. THE STM THEN PERFORMED PREVENTATIVE MAINTENANCE (PM) INCLUDING ALL SAFETY, FUNCTIONALITY, AND CALIBRATION CHECKS AND ALL TESTS PASSED TO FACTORY SPECIFICATIONS. THE IABP UNIT WAS CLEARED FOR CLINICAL USE AND RELEASED TO THE CUSTOMER. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA (B)(4).
IT WAS REPORTED THAT DURING USE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD BLOOD BACK INTO THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 453526 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |