FDA Adverse Event Malfunction Summary report: N

EASYSTAND EVOLV

MDR report key: 16062714 · Received December 28, 2022

Report

Report Number
2183634-2022-00003
Event Type
Malfunction
Date Received
December 28, 2022
Date of Event
October 15, 2022
Report Date
December 1, 2022
Manufacturer
ALTIMATE MEDICAL, INC.
Product Code
INO
PMA / PMN Number
K062402
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD FOR THE STANDER WAS REVIEWED. THE DEVICE IS MORE THAN 2 YEARS PAST ITS ESTIMATED LIFE. THERE HAVE BEEN NO ISSUES/COMPLAINTS PREVIOUSLY RECORDED FOR THE DEVICE SINCE ITS MANUFACTURE IN 2015. INCLUDED IN THE SAFETY PRECAUTIONS SECTION OF THE PRODUCTS OWNERS MANUAL THAT IS SENT WITH EVERY DEVICE STATES: IF THE EASYSTAND IS USED IN THE HOME ENVIRONMENT, PLEASE INSPECT THE UNIT PRIOR TO EACH USE TO ENSURE THERE IS NO DAMAGE OR UNEXPECTED WEAR TO THE EASYSTAND THAT MAY HAVE BEEN CAUSED UNINTENTIONALLY BY PETS, PESTS, AND CHILDREN. ALSO STATED: THE EXPECTED SERVICE LIFE OF AN EASYSTAND IS CONSIDERED 5 YEARS UNDER NORMAL USE CONDITIONS. IN ADDITION THERE IS A HYDRAULIC ACTUATOR MAINTENANCE SECTION IN THE PRODUCTS OWNERS MANUAL WHICH STATES THE FOLLOWING: HYDRAULIC ACTUATOR MAINTENANCE: IF THE PUMP IS LEAKING OIL OR NOT HOLDING PROPERLY CONTACT YOUR LOCAL EASYSTAND SUPPLIER FOR A REPLACEMENT, THE SERIAL NUMBER ON THE EASYSTAND WILL BE NEEDED. DO NOT EXERT EXCESSIVE FORCE ON THE PUMP HANDLE OR USE THE PUMP HANDLE TO MOVE AN EASYSTAND THIS MAY CAUSE DAMAGE TO THE PUMP HANDLE AND/OR THE PUMP MECHANISM AND WILL VOID ANY WARRANTY. USER MAINTENANCE IS NOT REQUIRED OR RECOMMENDED ON THE HYDRAULIC PUMP. NEVER APPLY LUBRICANT, HOUSEHOLD OIL, GREASE, OR A SIMILAR PRODUCT TO ANY PART OF THE UNIT. SOME COMPOUNDS MAY REACT WITH THE SEALS USED IN THE MANUFACTURE OF THIS PRODUCT. THIS MAY DAMAGE THE UNIT AND WILL VOID ANY WARRANTY. THE DISTRIBUTOR HAS REPLACED THE UNIT FOR THE CUSTOMER WITH A NEW STANDING AID.

Description of Event or Problem · 0

ON 12/1/2022, ALTIMATE MEDICAL RECEIVED A NOTICE FROM THE DISTRIBUTOR IN SWEDEN THERE WAS AN ISSUE REPORTED TO THE SWEDISH AUTHORITY. THE REPORT WAS SENT TO ALTIMATE MEDICAL IN SWEDISH AND GOOGLE TRANSLATE WAS USED TO TRANSLATE THE REPORT. THE REPORT ENTRY TRANSLATED AS: "SEQUENCE OF EVENTS - THE PATIENT WAS GOING TO STAND EXERCISE IN HIS EASYSTAND, WITH THE HELP OF AN ASSISTANT, WHEN THE PISTON BROKE (ACCORDING TO THE DESCRIPTION "EXPLODED") AND OIL LEAKED OUT. FORTUNATELY, NO ONE WAS PHYSICALLY HARMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2210589 EASYSTAND EVOLV STANDING FRAME INO ALTIMATE MEDICAL, INC. EVOLV LARGE

Patients

Seq Age Sex Outcome Treatment
1 Unknown