FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 16052775 · Received December 27, 2022

Report

Report Number
2024168-2022-13110
Event Type
Injury
Date Received
December 27, 2022
Date of Event
November 10, 2022
Report Date
February 18, 2023
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648013089
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT ESTIMATED. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 0

VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE UNSPECIFIED FAILURE WAS OBSERVED AS A NEEDLE TO CUFF MISS. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE NEEDLE TO CUFF MISS AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. D4: LOT # CHANGED FROM 2092841 TO 2091242.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VESSEL CLOSURE OF AN UNKNOWN VESSEL USING A PROGLIDE DEVICE RELATIVE TO AN UNKNOWN PROCEDURE. REPORTEDLY, AN UNKNOWN FAILURE WAS NOTICED. AN UNKNOWN METHOD WAS USED TO ACHIEVE HEMOSTASIS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2403736 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 12673-03 2091242 08717648013089

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention