FDA Adverse Event
Malfunction
Summary report: N
MP30 INTELLIVUE PATIENT MONITOR
MDR report key: 1604566
·
Received November 12, 2009
Report
- Report Number
- 9610816-2009-00202
- Event Type
- Malfunction
- Date Received
- November 12, 2009
- Report Date
- November 11, 2009
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MHX
- PMA / PMN Number
- K03003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER REPORTED THAT WHEN A MEMBER OF THE CLEANING STAFF MOVED THE OVERHEAD ARM TO WHICH THE MP30 MONITOR/GCX MOUNT WAS ATTACHED, THE GCX MOUNT FAILED, AND THE MONITOR FELL TO THE FLOOR. THE PICTURES OF THE FAILED MOUNT INDICATE METAL SHEARING FROM FORCE OUTSIDE NORMAL AND EXPECTED USE. PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETED. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHEN A MEMBER OF THE CLEANING STAFF MOVED THE OVERHEAD ARM TO WHICH THE MP30 MONITOR/GCX MOUNT WAS ATTACHED, THE GCX MOUNT FAILED, AND THE MONITOR FELL TO THE FLOOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MP30 INTELLIVUE PATIENT MONITOR | MHX | PHILIPS MEDICAL SYSTEMS | M8002A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |