FDA Adverse Event
Malfunction
Summary report: N
ELECSYS HIV DUO REAGENT PACK
MDR report key: 16043481
·
Received December 22, 2022
Report
- Report Number
- MW5113985
- Event Type
- Malfunction
- Date Received
- December 22, 2022
- Date of Event
- November 16, 2022
- Report Date
- November 28, 2022
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC.
- Product Code
- MZF
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
LAB SPECIMEN DRAWN FROM PT ON (B)(6) 2022. ELECSYS HIV DUO (AB/AG) PERFORMED ON ROCHE COBAS PRO PRODUCED A REPEATEDLY POSITIVE RESULT ON (B)(6) 2022. RESULT WAS REPORTED TO PROVIDER AND STATE AGENCY (SCDHEC). CONFIRMATORY TESTING ON THE SAME SPECIMEN SENT TO LABCORP FOR 4TH GENERATION HIV AB/AG. CONFIRMATORY TESTING WAS PERFORMED (B)(6) 2022 AND WAS NEGATIVE FOR HIV-1/HIV-2 AB AND NEGATIVE FOR P24 AG INDICATING NO LAB EVIDENCE OF HIV INFECTION. ROCHE COBAS PRO ELECSYS HIV DUO (AHIV AND HIVAG) AND ACCEPTABLE CALIBRATION AND QC FOR DATE OF TESTING. TESTING DISCREPANCY REPORTED TO ROCHE DIAGNOSTICS CASE # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2653875 | ELECSYS HIV DUO REAGENT PACK | TEST, HIV DETECTION | MZF | ROCHE MOLECULAR SYSTEMS, INC. | 582002 | ||
| 2653876 | ROCHE COBAS PRO | CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE | MZF | ROCHE MOLECULAR SYSTEMS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Female | Other |