FDA Adverse Event Malfunction Summary report: N

ELECSYS HIV DUO REAGENT PACK

MDR report key: 16043481 · Received December 22, 2022

Report

Report Number
MW5113985
Event Type
Malfunction
Date Received
December 22, 2022
Date of Event
November 16, 2022
Report Date
November 28, 2022
Manufacturer
ROCHE MOLECULAR SYSTEMS, INC.
Product Code
MZF
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

LAB SPECIMEN DRAWN FROM PT ON (B)(6) 2022. ELECSYS HIV DUO (AB/AG) PERFORMED ON ROCHE COBAS PRO PRODUCED A REPEATEDLY POSITIVE RESULT ON (B)(6) 2022. RESULT WAS REPORTED TO PROVIDER AND STATE AGENCY (SCDHEC). CONFIRMATORY TESTING ON THE SAME SPECIMEN SENT TO LABCORP FOR 4TH GENERATION HIV AB/AG. CONFIRMATORY TESTING WAS PERFORMED (B)(6) 2022 AND WAS NEGATIVE FOR HIV-1/HIV-2 AB AND NEGATIVE FOR P24 AG INDICATING NO LAB EVIDENCE OF HIV INFECTION. ROCHE COBAS PRO ELECSYS HIV DUO (AHIV AND HIVAG) AND ACCEPTABLE CALIBRATION AND QC FOR DATE OF TESTING. TESTING DISCREPANCY REPORTED TO ROCHE DIAGNOSTICS CASE # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2653875 ELECSYS HIV DUO REAGENT PACK TEST, HIV DETECTION MZF ROCHE MOLECULAR SYSTEMS, INC. 582002
2653876 ROCHE COBAS PRO CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE MZF ROCHE MOLECULAR SYSTEMS, INC.

Patients

Seq Age Sex Outcome Treatment
1 28 YR Female Other