FDA Adverse Event Malfunction Summary report: N

LEKSELL VANTAGE STEREOTACTIC SYSTEM

MDR report key: 16041678 · Received December 23, 2022

Report

Report Number
9612186-2022-00003
Event Type
Malfunction
Date Received
December 23, 2022
Date of Event
December 12, 2022
Report Date
March 10, 2023
Manufacturer
ELEKTA INSTRUMENT AB
Product Code
HAW
UDI-DI
07340048304399
PMA / PMN Number
K190887
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D8 UPDATED. H6 UPDATED. H10 UPDATED: THE ISSUE REPORTED BY THE CUSTOMER WAS THAT THE LEKSELL VANTAGE STEREOTACTIC SYSTEM HAD A TARGET SHIFT OF MORE THAN 2MM. THIS SITE IS USING A MICRODRIVE SUPPLIED BY A THIRD PARTY (ALPHA OMEGA) IN CONJUNCTION WITH THE LEKSELL VANTAGE STEREOTACTIC SYSTEM. THE LEKSELL VANTAGE STEREOTACTIC SYSTEM HAS BEEN INVESTIGATED REGARDING MECHANICAL ACCURACY AND IT IS WELL WITHIN SPECIFICATION. ELEKTA INSTRUMENT AB HAVE ALSO INVESTIGATED THE USED MICRODRIVE (ALPHA OMEGA) TOGETHER WITH THE LEKSELL VANTAGE STEREOTACTIC SYSTEM AND THE INVESTIGATION CONCLUDED THAT MICRODRIVE (ALPHA OMEGA) IS INCORRECTLY INTERFACING THE LEKSELL VANTAGE STEREOTACTIC SYSTEM IN A WAY THAT IT MAY CAUSE A SIGNIFICANT SHIFT BOTH ANTERIOR AND LATERAL (ACCURACY ERROR). THIS SHIFT IS SEEN FOR THE SYSTEM TESTED ONLY AND IS JUDGED TO BE HIGHLY DEPENDENT ON THE COMBINATION OF MICRODRIVE (ALPHA OMEGA) MEASUREMENTS AND THE STOP HOLDER MEASUREMENTS. THE RESULT WAS PRESENTED TO THE USER WHO CONFIRMS THAT THE DIRECTION IS THE SAME AS THEY HAVE SEEN DURING THEIR PROCEDURES AS WELL AS DURING PHANTOM TESTS. TESTS WERE ALSO REPEATED WITH THE LEKSELL VANTAGE STEREOTACTIC SYSTEM USING ELEKTA MICRODRIVE WHICH SHOW MUCH BETTER ACCURACY (WITHIN SPECIFICATION) THAN USING THE ALPHA OMEGA MICRODRIVE DUE TO CORRECT INTERFACING TO THE SYSTEM. THE ROOT CAUSE COULD THEREFORE BE CONCLUDED TO BE CAUSED BY INCOMPATIBILITY BY A THIRD PARTY PRODUCT. ALPHA OMEGA IS CLAIMING COMPATIBILITY WITH ALL ELEKTA INSTRUMENT AB'S STEREOTACTIC SYSTEMS ON THE MARKET. ELEKTA INSTRUMENT AB HAS ON SEVERAL OCCASIONS INFORMED ALPHA OMEGA THAT INTERFACING TOWARDS ONLY THE STOP HOLDER IS INCORRECT AND WILL CAUSE ACCURACY ISSUES. ELEKTA INSTRUMENT AB RECOMMENDED ALPHA OMEGA TO REVIEW THEIR INTERFACING DESIGN AND STATED THAT ALPHA OMEGA CANNOT CLAIM COMPATIBILITY WITH ELEKTA INSTRUMENT AB'S STEREOTACTIC SYSTEMS. LEKSELL VANTAGE STEREOTACTIC SYSTEM INHIBITED SAFELY AND THERE HAVE BEEN NO PATIENT INJURIES. PLEASE NOTE THAT THERE ARE TWO CASES REPORTED FROM THE SAME USER USING THE SAME SYSTEM SINCE IT WAS USED TWICE WHERE A DISCREPANCY WAS SEEN (ELEKTA REF. #(B)(4) - MFR NO 9612186-2022-00004).

Additional Manufacturer Narrative · 0

THE MANUFACTURER'S INVESTIGATION IS ON-GOING AND FURTHER INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED ACCURACY DEVIATIONS WITH LEKSELL VANTAGE DURING SURGERY WITH A SYSTEMATIC ERROR OF ABOUT 1.6 MM. THE CUSTOMER WAS USING THE NON-COMPATIBLE MICRODRIVE FROM THE MANUFACTURER ALPHA-OMEGA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2782792 LEKSELL VANTAGE STEREOTACTIC SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW ELEKTA INSTRUMENT AB 1053958 07340048304399

Patients

Seq Age Sex Outcome Treatment
1 Unknown