FDA Adverse Event Malfunction Summary report: N

LAPRA-TY ABSORBABLE X6::SUTURE CLIP

MDR report key: 16040310 · Received December 22, 2022

Report

Report Number
2210968-2022-10592
Event Type
Malfunction
Date Received
December 22, 2022
Date of Event
November 17, 2022
Report Date
January 17, 2023
Manufacturer
ETHICON INC.
Product Code
FZP
UDI-DI
10705031206885
PMA / PMN Number
K931492
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION WAS REQUESTED, AND THE FOLLOWING WAS OBTAINED VIA: PLEASE PROVIDE THE APPLIER PRODUCT CODE AND LOT NUMBER? KA200(2013-036 (01)10705036012641). WAS THE APPLIER CHECKED FOR DAMAGED (JAWS STRAIGHT AND ALIGNED)? VISUAL CHECK OF THE DEVICE BY SALES REP WAS WITH NO APPARENT DAMAGE. IF THE CLIP DID NOT CLOSE/HOLD ON THE SUTURE, WAS THE CLIP USED IN AN APPLICATION WHERE HE SUTURE WAS UNDER TENSION? THE SUTURE WAS UNDER TENSION TO SOME EXTENT. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 1/17/2023. H6 COMPONENT CODE: G07002 NO DEVICE PROBLEM FOUND. H3 INVESTIGATIONAL SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION WAS CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLES REVEALED THAT XC200 RELOAD WAS RECEIVED WITH NO APPARENT DAMAGE AND TWO CLIPS LOADED. THE RELOAD WAS TESTED FOR FUNCTIONALITY WITH THE TEST DEVICE. UPON FUNCTIONAL TESTING OF THE CLIPS, THE INSTRUMENT LOADED, RETAINED, AND DEPLOYED TWO CLIPS AS INTENDED. THE CLIPS WERE AS INTENDED AND CONFORMS TO OUR MANUFACTURING REQUIREMENTS. AS PART OF ETHICON¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE CLIP PERFORMED WITHOUT ANY DIFFICULTIES NOTED. ALTHOUGH NO PRODUCT DEFECT WAS IDENTIFIED, THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT COULD NOT BE REPLICATED DURING THE LABORATORY ANALYSIS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SPLENECTOMY PROCEDURE ON (B)(6) 2022 AND CLIPS SUTURE WERE USED. DURING THE PROCEDURE, THE DEVICE COULD NOT BE RATCHETTED ALTHOUGH THE TRIGGER WAS GRASPED. THE DEVICE WAS USED ON THE VICRYL 4-0. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2110115 LAPRA-TY ABSORBABLE X6::SUTURE CLIP CLIP, IMPLANTABLE FZP ETHICON INC. XC200 RM2AAT 10705031206885

Patients

Seq Age Sex Outcome Treatment
1 Unknown