FDA Adverse Event Malfunction Summary report: N

BD PHASEAL¿ SPIKE CONNECTOR (C180J)

MDR report key: 16030310 · Received December 21, 2022

Report

Report Number
3003152976-2022-00579
Event Type
Malfunction
Date Received
December 21, 2022
Date of Event
December 2, 2022
Report Date
February 1, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
ONB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY ONE SAMPLE AND THREE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, THE SPIKE IS OBSERVED TO BE BROKEN OFF OF THE UPPER BASED. FURTHER EVALUATION DID NOT IDENTIFY ANY AIR BUBBLES OR OTHER DEFECTS THAT WOULD HAVE LED TO THE BREAKAGE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2201210, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. PRODUCT UNDERGOES INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS TO ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE, NO ISSUES WERE IDENTIFIED. BASED ON THE SAMPLE EVALUATION, THE ROOT CAUSE WAS DETERMINED TO BE THE FORCE USED WHEN INSERTING THE SPIKE CONNECTOR INTO THE IV BAG.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PHASEAL¿ SPIKE CONNECTOR (C180J) EXPERIENCED A BROKEN TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE NEEDLE (SPIKE) OF 180J BROKE OFF FROM ITS BASE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PHASEAL¿ SPIKE CONNECTOR (C180J) EXPERIENCED A BROKEN TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE NEEDLE (SPIKE) OF 180J BROKE OFF FROM ITS BASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2858034 BD PHASEAL¿ SPIKE CONNECTOR (C180J) INTRAVASCULAR ADMINISTRATION SET ONB BECTON DICKINSON, S.A. 2201210

Patients

Seq Age Sex Outcome Treatment
1 Unknown