FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 16023721 · Received December 21, 2022

Report

Report Number
3001421318-2022-00094
Event Type
Malfunction
Date Received
December 21, 2022
Date of Event
December 14, 2022
Report Date
March 31, 2022
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002800747
PMA / PMN Number
K181216
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG HAS NOT RECEIVED THE DEVICE FOR INVESTIGATION SO FAR. HAMILTON MEDICAL AG CASE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

THE INCIDENT WAS REPORTED TO THE MANUFACTURER "PHILIPS AG" FOR INVESTIGATION. ACCORDING TO THEIR INVESTIGATION, THIS IS THE FIRST COMPLAINT OF THIS KIND THAT WAS RECEIVED SO FAR. THE DEVICE HISTORY RECORD SHOWED THAT THE DEVICE WAS MANUFACTURED ACCORDING TO SPECIFICATIONS. FURTHERMORE, THE FIXATION WAS REVIEWED AND IS FOUND TO BE SUFFICIENT. ON THE BASIS OF THE INFORMATION WE HAVE COLLECTED, WE ASSUME THAT THERE IS AN ISOLATED CASE IN THE HOSPITAL IN SWITZERLAND IN WHICH UNFAVORABLE INFLUENCES HAVE ACCUMULATED THAT WE HAVE NOT BEEN ABLE TO DETERMINE, REPRODUCE AND EVALUATE IN A QUALIFIED MANNER TO DATE. HAMILTON MEDICAL AG CASE NUMBER IS: (B)(4).

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: UNIVERSAL BED MOUNT 161148 ALMOST FELL ON A PATIENT, SHE FIXED THE EQUIPMENT. THE C1 IS MOUNTED ABOVE THE HEAD OF THE PATIENT AND THAT WOULD BE A REAL PITY IF IT FALLS ON THE HEAD. NO PATIENT HARM OCCURRED AS VENTILATOR WAS NEVER ON PATIENT.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: UNIVERSAL BED MOUNT 161148 ALMOST FELL ON A PATIENT, SHE FIXED THE EQUIPMENT. THE C1 IS MOUNTED ABOVE THE HEAD OF THE PATIENT AND THAT WOULD BE A REAL PITY IF IT FALLS ON THE HEAD. NO PATIENT HARM OCCURRED AS VENTILATOR WAS NEVER ON PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476266 HAMILTON MEDICAL AG HAMILTON-C1 CBK HAMILTON MEDICAL AG HAMILTON-C1 07630002800747

Patients

Seq Age Sex Outcome Treatment
1 Unknown