FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ INSULIN SYRINGE

MDR report key: 16018683 · Received December 20, 2022

Report

Report Number
1920898-2022-00894
Event Type
Malfunction
Date Received
December 20, 2022
Date of Event
November 20, 2022
Report Date
January 9, 2023
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1221371. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATION.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ INSULIN SYRINGE WAS BLOCKED AND COULDN'T DELIVER INSULIN DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "ON (B)(6) 2022, DUE TO DIABETES, THE PATIENT NEEDED TO BE GIVEN HUMAN INSULIN INJECTION HYPOGLYCEMIC TREATMENT BEFORE MEALS. DURING THE PROCESS OF PUMPING INSULIN, INSULIN COULD NOT BE PUMPED OUT WITH AN INSULIN SYRINGE, SO IT WAS REPLACED IMMEDIATELY."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ INSULIN SYRINGE WAS BLOCKED AND COULDN'T DELIVER INSULIN DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "ON (B)(6) 2022, DUE TO DIABETES, THE PATIENT NEEDED TO BE GIVEN HUMAN INSULIN INJECTION HYPOGLYCEMIC TREATMENT BEFORE MEALS. DURING THE PROCESS OF PUMPING INSULIN, INSULIN COULD NOT BE PUMPED OUT WITH AN INSULIN SYRINGE, SO IT WAS REPLACED IMMEDIATELY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2052206 BD ULTRA-FINE¿ INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 1221371

Patients

Seq Age Sex Outcome Treatment
1 Unknown