FDA Adverse Event Malfunction Summary report: N

BD PHASEAL¿ SECONDARY SET C61

MDR report key: 16006043 · Received December 19, 2022

Report

Report Number
2243072-2022-02177
Event Type
Malfunction
Date Received
December 19, 2022
Date of Event
November 29, 2022
Report Date
January 15, 2023
Manufacturer
BECTON DICKINSON
Product Code
FPA
UDI-DI
00382905153020
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1022536. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, A PHOTOGRAPH WAS SUBMITTED TO OUR FACILITY, WHICH DISPLAYED THE REPORTED LEAKAGE OCCURRING BETWEEN THE TUBE AND SPIKE. THE ISSUE HAS BEEN CONFIRMED. OUR ENGINEERS HAVE FURTHER IDENTIFY THE MOST LIKELY ROOT CAUSE AS TOO LITTLE SOLVENT ON THE JOINT, WHICH DIRECTLY LEAD TO THE LEAK. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

INITIAL REPORTER ADDR 1: (B)(6). DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(6). THIS SITE IS NOT REGISTERED WITH THE FDA. (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PHASEAL¿ SECONDARY SET C61 LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE DURING INFUSION CAUSING SPILLAGE IN THE WARDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PHASEAL¿ SECONDARY SET C61 LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE DURING INFUSION CAUSING SPILLAGE IN THE WARDS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2126496 BD PHASEAL¿ SECONDARY SET C61 INTRAVASCULAR ADMINISTRATION SET FPA BECTON DICKINSON 1022536 00382905153020

Patients

Seq Age Sex Outcome Treatment