BD PHASEAL¿ SECONDARY SET C61
Report
- Report Number
- 2243072-2022-02177
- Event Type
- Malfunction
- Date Received
- December 19, 2022
- Date of Event
- November 29, 2022
- Report Date
- January 15, 2023
- Manufacturer
- BECTON DICKINSON
- Product Code
- FPA
- UDI-DI
- 00382905153020
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1022536. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, A PHOTOGRAPH WAS SUBMITTED TO OUR FACILITY, WHICH DISPLAYED THE REPORTED LEAKAGE OCCURRING BETWEEN THE TUBE AND SPIKE. THE ISSUE HAS BEEN CONFIRMED. OUR ENGINEERS HAVE FURTHER IDENTIFY THE MOST LIKELY ROOT CAUSE AS TOO LITTLE SOLVENT ON THE JOINT, WHICH DIRECTLY LEAD TO THE LEAK. H3 OTHER TEXT : SEE H10.
INITIAL REPORTER ADDR 1: (B)(6). DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(6). THIS SITE IS NOT REGISTERED WITH THE FDA. (B)(4).
IT WAS REPORTED THAT THE BD PHASEAL¿ SECONDARY SET C61 LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE DURING INFUSION CAUSING SPILLAGE IN THE WARDS.
IT WAS REPORTED THAT THE BD PHASEAL¿ SECONDARY SET C61 LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKAGE DURING INFUSION CAUSING SPILLAGE IN THE WARDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2126496 | BD PHASEAL¿ SECONDARY SET C61 | INTRAVASCULAR ADMINISTRATION SET | FPA | BECTON DICKINSON | 1022536 | 00382905153020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|