FDA Adverse Event Malfunction Summary report: N

ARCHITECT C16000

MDR report key: 16001630 · Received December 18, 2022

Report

Report Number
3016438761-2022-00592
Event Type
Malfunction
Date Received
December 18, 2022
Date of Event
November 25, 2022
Report Date
January 20, 2023
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740005924
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL REPORT WAS SUBMITTED UNDER MANUFACTURER REPORT NUMBER 3002809144-2022-00436 ((B)(4), GERMANY AS MANUFACTURING SITE) WITH SUSPECT MEDICAL DEVICE OF ARCHITECT TOTAL BILIRUBIN, (B)(4). AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED TO ARCHITECT C16000, (B)(4) ((B)(4), TX AS MANUFACTURING SITE), WHICH IS THIS REPORT. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 0

FIELD SERVICE IDENTIFIED AND DETERMINED THE AERO CUV PR DRY THE LIKELY CAUSE FOR THE FALSELY LOW TOTAL BILIRUBIN RESULTS AS THE MODULES CUVETTE SHOWED SIGNS OF BEING DIRTY/CONTAMINATED. THE AERO CUV PR DRY WAS CLEANED TO RESOLVE THE ISSUE. THE MODULE FUNCTION WAS VERIFIED WITH A CONTROL RUN. LABELING REVIEW CONCLUDES THAT THE ISSUE IS ADEQUATELY ADDRESSED. A REVIEW OF THE ARCHITECT C16000, SERIAL (B)(6) SERVICE HISTORY, REVEALED NO ADDITIONAL ISSUES OF FALSE LOW TOTAL BILIRUBIN RESULTS REPORTED PRE OR POST FIELD SERVICE INTERVENTION FOR THE CURRENT TICKET. TRENDING REVIEW DID NOT IDENTIFY ANY TRENDS FOR THE ISSUE FOR THE PRODUCT. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES OR POTENTIAL NONCONFORMANCES APPLICABLE TO THE COMPLAINT ISSUE. NO PRODUCT DEFICIENCY WAS IDENTIFIED. SECTION B5, EVENT DESCRIPTION UPDATED FROM FALSE ELEVATED ARCHITECT TOTAL BILIRUBIN RESULTS THAT REPEATED LOWER AS STATED ON THE INITIAL REPORT TO FALSE DEPRESSED ARCHITECT TOTAL BILIRUBIN RESULTS THAT REPEATED HIGHER ON THE THIS CORRECTED REPORT. SECTION H6, MEDICAL DEVICE PROBLEM CODE UPDATED FROM A090809 HIGH TEST ON THE INITIAL REPORT TO A090810 LOW TEST ON THIS CORRECTED REPORT.

Description of Event or Problem · 0

THE ACCOUNT GENERATED FALSE ELEVATED ARCHITECT TOTAL BILIRUBIN RESULTS ON 4 SAMPLES THAT REPEATED LOWER WHEN PROCESSING ON THE ARCHITECT C16000. THE ACCOUNT USES A REFERENCE RANGE OF 3.4 TO 20.5 UMOL/L. (B)(6) (FEMALE, 32 YEAR OLD, HAN/CHINESE) TESTED 1.7 REPEATED 7.2 UMOL/L. (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 1.40 REPEATED 12.4 UMOL/L. (B)(6) (FEMALE, 28 YEAR OLD, HAN/CHINESE) TESTED 3.1 REPEATED 8.1 UMOL/L. (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 0.8 REPEATED 7.2 UMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE ACCOUNT GENERATED FALSE DEPRESSED ARCHITECT TOTAL BILIRUBIN RESULTS ON 4 SAMPLES THAT REPEATED HIGHER WHEN PROCESSING ON THE ARCHITECT C16000. THE ACCOUNT USES A REFERENCE RANGE OF 3.4 TO 20.5 UMOL/L. SID (B)(6) (FEMALE, 32 YEAR OLD, HAN/CHINESE) TESTED 1.7 REPEATED 7.2 UMOL/L. SID (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 1.40 REPEATED 12.4 UMOL/L. SID (B)(6) (FEMALE, 28 YEAR OLD, HAN/CHINESE) TESTED 3.1 REPEATED 8.1 UMOL/L. SID (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 0.8 REPEATED 7.2 UMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2152426 ARCHITECT C16000 ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 3L77-01 00380740005924

Patients

Seq Age Sex Outcome Treatment
1 Female CC TOTAL BILI (2750T), 06L45-22, UNKNOWN.| CC TOTAL BILI (2750T), 06L45-22, UNKNOWN.