HOYA 255
Report
- Report Number
- 3006723646-2022-00159
- Event Type
- Injury
- Date Received
- December 16, 2022
- Date of Event
- November 8, 2022
- Report Date
- November 17, 2022
- Manufacturer
- HOYA SURGICAL OPTICS, INC.
- Product Code
- HQL
- PMA / PMN Number
- P080004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THIS INITIAL REPORT IS BEING SUBMITTED TO FDA FOR A REPORTABLE EVENT THAT OCCURRED OUTSIDE THE USA. HAPTIC DAMAGE IS INDICATED AS A POTENTIAL MALFUNCTION RELATED TO THE IOL, AS COVERED UNDER THE WARNINGS SECTION OF THE PRODUCT'S INSTRUCTIONS FOR USE (IFU). PORTIONS OF THE PRODUCT WERE RETURNED TO THE MANUFACTURER. THE INVESTIGATION WAS CONDUCTED, WITH THE METHODS AND RESULTS AS NOTED BELOW. APPEARANCE CHECK RESULT WAS CONSISTENT WITH THE REPORTED INFORMATION. NO ABNORMALITIES WERE FOUND IN PRODUCTION AND INSPECTION RECORDS OF THE PRODUCT. (SERIAL NO.: (B)(4); MODEL: 255). EXACT ROOT CAUSE WAS NOT DETERMINED. HOWEVER, BASED ON THE AVAILABLE INFORMATION AND OUR INVESTIGATION, WE BELIEVE THIS EVENT WAS NOT CAUSED BY OUR PRODUCT QUALITY. CAPA-22-0009 HAS BEEN INITIATED FOR "DAMAGED HAPTIC" COMPLAINTS.
DAMAGED HAPTIC AFTER IMPLANTATION. THE DOCTOR CUT AND EXPLANTED THE IOL FROM THE EYE SINCE HE/SHE NOTICED THE TRAILING HAPTIC WAS SEPARATED AND THE OPTIC GOT CRACKED. PATIENT IMPACT: INTRA-OPERATIVE EXPLANTATION
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2747723 | HOYA 255 | INTRAOCULAR LENS | HQL | HOYA SURGICAL OPTICS, INC. | HOYA 255 (+16.50 D) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |