FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOGUARD PNK 20GA X 1.16IN

MDR report key: 15996940 · Received December 16, 2022

Report

Report Number
1710034-2022-00817
Event Type
Malfunction
Date Received
December 16, 2022
Date of Event
November 23, 2022
Report Date
February 21, 2023
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903818341
PMA / PMN Number
K952861
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: OUR QUALITY ENGINEER INSPECTED THE PHOTOGRAPHS SUBMITTED FOR EVALUATION. BD RECEIVED TWO PHOTOS OF 20GA X 1.16IN INSYTE AUTOGUARD UNITS. THE PHOTOS DISPLAY THE UNITS IN A BOX FROM LOT NUMBER 2112426. BASED ON THE DESCRIPTION OF YOUR REPORT, THESE ARE THE UNITS THAT GOT MIXED WITH PRODUCT FROM LOT NUMBER 2109740. THE MATERIAL NUMBER IS THE SAME. THE REPORTED ISSUE WAS CONFIRMED. MIXED LOTS MAY OCCUR DURING MANUFACTURING OR PACKAGING DUE TO IMPROPER LINE CLEARANCE BY OPERATORS OR AN IMPROPER LINE PURGE DUE TO EQUIPMENT SOFTWARE ERROR. OPERATORS ALSO PERFORM PERIODIC VISUAL INSPECTIONS PER THE QUALITY PLAN TO MITIGATE THE OCCURRENCE OF THIS DEFECT. THE DISTRIBUTION CENTER WAS CONTACTED TO FIND WHERE THE LOTS MAY HAVE BEEN MIXED AT. THE DISTRIBUTION LIST RECORDS THAT "DELPHARM EVREUX" WAS ONLY SUPPOSED TO RECEIVE UNITS FROM LOT NUMBER 2109740. THERE IS NO RECORD OF LOT NUMBER 2112426 BEING SHIPPED TO THEM. SINCE THERE IS PHOTO EVIDENCE OF THE LOT NUMBER 2112426 BEING PRESENT IN THEIR FACILITY THIS DEFECT IS CONFIRMED. THE MIXED PRODUCT LIKELY OCCURRED WITHIN BD CONTROL AND MAY HAVE OCCURRED DURING PACKAGING OR IN THE DISTRIBUTION CENTER. THERE WAS NO EVIDENCE OF MISLABELING OR THAT LOTS WERE MIXED WITHIN THE SHIPPER BOXES. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THESE BATCHES.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD INSYTE AUTOGUARD PNK 20GA X 1.16IN THERE WAS A MIXTURE OF PRODUCTS IN ONE PACK. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING AN IN-PROCESS CONTROL IN PRODUCTION (IPC EVERY 30 MINUTES) ON (B)(6) 2022 THE OPERATOR FOUND A MIX-UP BETWEEN BATCH 2109740 (SUPPLIER'S BATCH) AND BATCH 2112426. SOME PRODUCT FROM BATCH 2112426 WERE FOUND INSIDE THE BATCH 2109740. 200 MIXED CATHETERS WERE FOUND DURING THE SORT. ADDITIONALLY, A PREVIOUS RECLAMATION CONCERNING A MIX-UP OF THIS PRODUCT WAS SENT THE (B)(6) 2021.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD INSYTE AUTOGUARD PNK 20GA X 1.16IN THERE WAS A MIXTURE OF PRODUCTS IN ONE PACK. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING AN IN-PROCESS CONTROL IN PRODUCTION (IPC EVERY 30 MINUTES) ON NOVEMBER 23RD, 2022 THE OPERATOR FOUND A MIX-UP BETWEEN BATCH 2109740 (SUPPLIER'S BATCH) AND BATCH 2112426. SOME PRODUCT FROM BATCH 2112426 WERE FOUND INSIDE THE BATCH 2109740. 200 MIXED CATHETERS WERE FOUND DURING THE SORT. ADDITIONALY, A PREVIOUS RECLAMATION CONCERNING A MIX-UP OF THIS PRODUCT WAS SENT ON THE 22/09/2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2201788 BD INSYTE AUTOGUARD PNK 20GA X 1.16IN INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 381834 2109740 00382903818341

Patients

Seq Age Sex Outcome Treatment
1 Unknown