BD ULTRA-FINE?¿ II INSULIN SYRINGE
Report
- Report Number
- 1920898-2022-00880
- Event Type
- Malfunction
- Date Received
- December 15, 2022
- Date of Event
- November 24, 2022
- Report Date
- January 10, 2023
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY CUSTOMER RETURNED AN IMAGE OF A SINGLE 0.3ML SYRINGE. THE SYRINGE HAD ITS NEEDLE SHIELD AND HUB SEPARATE FROM THE BARREL. THE HUB HAS BECOME LODGED INSIDE THE SHIELD. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1123322. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE PHOTO RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. H3 OTHER TEXT : SEE H10
IT WAS REPORTED THAT THE NEEDLE HUB SEPARATED FROM THE BD ULTRA-FINE¿ II INSULIN SYRINGE WHEN REMOVING THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "A PET CAREGIVER GETS IN TOUCH TO REPORT THAT ONE OF THE 30UI BD ULTRA-FINE SYRINGES HAD THE FOLLOWING QUALITY DEVIATION: WHEN REMOVING THE ORANGE PROTECTIVE COVER, THE NEEDLE DETACHED ITSELF FROM THE SYRINGE BODY AND REMAINED ATTACHED TO THE PROTECTIVE COVER."
IT WAS REPORTED THAT THE NEEDLE HUB SEPARATED FROM THE BD ULTRA-FINE¿ II INSULIN SYRINGE WHEN REMOVING THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "A PET CAREGIVER GETS IN TOUCH TO REPORT THAT ONE OF THE 30UI BD ULTRA-FINE SYRINGES HAD THE FOLLOWING QUALITY DEVIATION: WHEN REMOVING THE ORANGE PROTECTIVE COVER, THE NEEDLE DETACHED ITSELF FROM THE SYRINGE BODY AND REMAINED ATTACHED TO THE PROTECTIVE COVER."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2734812 | BD ULTRA-FINE?¿ II INSULIN SYRINGE | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 1123322 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |