Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY CUSTOMER RETURNED AN IMAGE OF A SINGLE 0.3ML SYRINGE. THE SYRINGE HAD ITS NEEDLE SHIELD AND HUB SEPARATE FROM THE BARREL. THE HUB HAS BECOME LODGED INSIDE THE SHIELD. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1123322. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE PHOTO RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. H3 OTHER TEXT : SEE H10

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE HUB SEPARATED FROM THE BD ULTRA-FINE¿ II INSULIN SYRINGE WHEN REMOVING THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "A PET CAREGIVER GETS IN TOUCH TO REPORT THAT ONE OF THE 30UI BD ULTRA-FINE SYRINGES HAD THE FOLLOWING QUALITY DEVIATION: WHEN REMOVING THE ORANGE PROTECTIVE COVER, THE NEEDLE DETACHED ITSELF FROM THE SYRINGE BODY AND REMAINED ATTACHED TO THE PROTECTIVE COVER."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE HUB SEPARATED FROM THE BD ULTRA-FINE¿ II INSULIN SYRINGE WHEN REMOVING THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "A PET CAREGIVER GETS IN TOUCH TO REPORT THAT ONE OF THE 30UI BD ULTRA-FINE SYRINGES HAD THE FOLLOWING QUALITY DEVIATION: WHEN REMOVING THE ORANGE PROTECTIVE COVER, THE NEEDLE DETACHED ITSELF FROM THE SYRINGE BODY AND REMAINED ATTACHED TO THE PROTECTIVE COVER."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2734812 BD ULTRA-FINE?¿ II INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 1123322

Patients

Seq Age Sex Outcome Treatment
1 Unknown