FDA Adverse Event Malfunction Summary report: N

VICRYL RAP UND 45CM 5/0 P-3 PRIME

MDR report key: 15988009 · Received December 15, 2022

Report

Report Number
2210968-2022-10307
Event Type
Malfunction
Date Received
December 15, 2022
Date of Event
January 1, 2022
Report Date
March 7, 2023
Manufacturer
ETHICON INC.
Product Code
GAM
PMA / PMN Number
K033746
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DATE SENT TO THE FDA: 12/15/2022. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED. H6 COMPONENT CODE: G07002 DEVICE NOT RETURNED ATTEMPTS ARE BEING MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION WAS REQUESTED, AND THE FOLLOWING WAS OBTAINED: WHAT DO YOU MEAN BY "WITH EASE"? PLEASE CLARIFY THE THREAD BREAKS EASILY EVEN WHEN THE NEEDLE HOLDER HAS BEEN CHANGED (4 DIFFERENT ONES HAVE BEEN USED) WHAT WAS BEING DONE WHEN THE SUTURE BROKE (REMOVAL FROM PACKAGE /DURING PASSAGE THROUGH TISSUE/ DURING TYING / POST-OP)? THE SURGEON REPORTS THE SUTURE'S THREAD BROKE WHILE HE WAS PULLING IT. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION RECEIVED STATED: "THE THREAD BREAKS EASILY EVEN WHEN THE NEEDLE HOLDER HAS BEEN CHANGED (4 DIFFERENT ONES HAVE BEEN USED)." PLEASE CONFIRM: DOES THE STATEMENT "4 DIFFERENT ONES" REFER TO THE NUMBER OF NEEDLE HOLDERS USED? HOW MANY SUTURES BROKE?

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 01/09/2023. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: THE STATEMENT "FOUR DIFFERENT ONES" REFERS TO THE NEEDLE HOLDERS. ONLY ONE THREAD BROKE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 3/7/2023. H6 COMPONENT CODE: G07002 NO DEVICE PROBLEM FOUND. ADDITIONAL INFORMATION: D9, H3, H6. H3 INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TEST EVALUATION WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT ONE OPENED BOX WITH THIRTY-SIX PACKETS THAT PERTAIN TO PRODUCT CODE V4930H WAS RECEIVED FOR EVALUATION. UPON INITIAL INSPECTION, OF THE SAMPLES, NO EXTERNAL DAMAGES WERE OBSERVED ON THE PACKETS. IN ORDER TO EVALUATE THE CONDITIONS OF THE RETURNED SAMPLES, THIRTEEN PACKETS WERE OPENED, AND NO DEFECTS WERE DETECTED. THE SWAGE AND ATTACHMENT AREA WERE NOTED TO BE AS EXPECTED. THE SUTURES WERE DISPENSED WITHOUT PROBLEMS AND EXAMINED ALONG OF THE STRAND TO DETECT ANY ISSUE RELATED TO DAMAGED OR BREAKAGE SUTURES AND NO DEFECTS WERE OBSERVED DURING EVALUATION. A FUNCTIONAL TEST WAS PERFORMED, AND THE TENSILE STRENGTH RESULTS WERE ABOVE THE MINIMUM REQUIREMENTS. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DEFECT NOTED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH BATCH IS RANDOMLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, A RUPTURE OF THE SUTURE WITH EASE OCCURRED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1756825 VICRYL RAP UND 45CM 5/0 P-3 PRIME SUTURE, ABSORBABLE, SYNTHETIC GAM ETHICON INC. SCBCLRT0

Patients

Seq Age Sex Outcome Treatment
1 Unknown