FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 1598700 · Received February 12, 2010

Report

Report Number
2250051-2010-00031
Event Type
Malfunction
Date Received
February 12, 2010
Date of Event
February 1, 2010
Report Date
February 12, 2010
Product Code
JTC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) FOUND DRIED SERUM ON THE TIP AND PLUNGER CLAMP IN POS #12 AND A DEFECTIVE PLUNGER CLAMP ON POS#1. FSE REPLACED BOTH THE PLUNGER CLAMPS AND THE TIP CLAMP. FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTER JTC 7003009

Patients

Seq Age Sex Outcome Treatment
1