FDA Adverse Event Malfunction Summary report: N

IMPACTOR HANDLE

MDR report key: 15983842 · Received December 14, 2022

Report

Report Number
1818910-2022-25471
Event Type
Malfunction
Date Received
December 14, 2022
Date of Event
December 1, 2022
Report Date
December 14, 2022
Manufacturer
DEPUY IRELAND - 3015516266
Product Code
LXH
UDI-DI
10603295539384
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT (B)(4). INVESTIGATION SUMMARY: THE DEVICE ASSOCIATED WITH THIS REPORT WAS RETURNED FOR ANALYSIS. VISUAL ANALYSIS OF THE DEVICE WAS ABLE TO CONFIRM, THE COMPLAINT. THE THREADED TIP OF THE ROD HAS STRIPPED. ADDITIONAL MONITORING, FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A SEARCH OF THE DEPUY NONCONFORMANCE (NC) QUALITY SYSTEM FOUND, NO NC¿S ASSOCIATED WITH THIS PRODUCT/LOT COMBINATION. BASED ON THE INABILITY TO FIND ANY NC¿S AGAINST THE PROVIDED PRODUCT CODE/LOT CODE COMBINATION, IT IS REASONABLE TO CONCLUDE, THAT THERE ARE NO ANOMALIES WITH REGARD TO MANUFACTURING OR INSPECTION CONTAINED IN THE DEVICE HISTORY RECORDS THAT WOULD CONTRIBUTE TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED STATES THAT THE THREADS APPEAR TO BE PEELING OFF. ALL PARTS ACCOUNTED FOR.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE THREADS ON IMPACTORS WERE DAMAGED. IT DID NOT HAPPENED DURING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1757552 IMPACTOR HANDLE SHOULDER INSTRUMENT - IMPACTOR HANDLE LXH DEPUY IRELAND - 3015516266 6205-10-101 PU167684 10603295539384

Patients

Seq Age Sex Outcome Treatment
1 Unknown CONCAVE IMPACTOR TIP| CONCAVE IMPACTOR TIP