FDA Adverse Event Malfunction Summary report: N

TOTAL BILIRUBIN

MDR report key: 15973138 · Received December 13, 2022

Report

Report Number
3002809144-2022-00436
Event Type
Malfunction
Date Received
December 13, 2022
Date of Event
November 25, 2022
Report Date
December 18, 2022
Manufacturer
ABBOTT GMBH
Product Code
CIG
UDI-DI
00380740174965
PMA / PMN Number
K121985
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED TO ARCHITECT C16000, LIST 3L77 (IRVING, TX AS MANUFACTURING SITE) AND SUBMITTED UNDER MANUFACTURER REPORT NUMBER 3016438761-2022-00592. ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER MDR NUMBER 3016438761-2022-00592.AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED TO ARCHITECT C16000, LIST 3L77 (IRVING, TX AS MANUFACTURING SITE) AND SUBMITTED UNDER MANUFACTURER REPORT NUMBER 3016438761-2022-00592. ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER MDR NUMBER 3016438761-2022-00592.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE ACCOUNT GENERATED FALSE ELEVATED ARCHITECT TOTAL BILIRUBIN RESULTS ON 4 SAMPLES THAT REPEATED LOWER WHEN PROCESSING ON THE ARCHITECT C16000. THE ACCOUNT USES A REFERENCE RANGE OF 3.4 TO 20.5 UMOL/L. SID (B)(6) (FEMALE, 32 YEAR OLD, HAN/CHINESE) TESTED 1.7 REPEATED 7.2 UMOL/L. SID (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 1.40 REPEATED 12.4 UMOL/L. SID (B)(6) (FEMALE, 28 YEAR OLD, HAN/CHINESE) TESTED 3.1 REPEATED 8.1 UMOL/L. SID (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 0.8 REPEATED 7.2 UMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE ACCOUNT GENERATED FALSE ELEVATED ARCHITECT TOTAL BILIRUBIN RESULTS ON 4 SAMPLES THAT REPEATED LOWER WHEN PROCESSING ON THE ARCHITECT C16000. THE ACCOUNT USES A REFERENCE RANGE OF 3.4 TO 20.5 UMOL/L. SID (B)(6) (FEMALE, 32 YEAR OLD, HAN/CHINESE) TESTED 1.7 REPEATED 7.2 UMOL/L. SID (B)(6) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 1.40 REPEATED 12.4 UMOL/L. SID (B)(6) (FEMALE, 28 YEAR OLD, HAN/CHINESE) TESTED 3.1 REPEATED 8.1 UMOL/L. SID (B)(4) (FEMALE, 29 YEAR OLD, HAN/CHINESE) TESTED 0.8 REPEATED 7.2 UMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1996111 TOTAL BILIRUBIN DIAZO COLORIMETRY, BILIRUBIN CIG ABBOTT GMBH 6L45-22 00380740174965

Patients

Seq Age Sex Outcome Treatment
1 Female ARC C16K PRC MOD, 03L77-01, (B)(6) | ARC C16K PRC MOD, 03L77-01, (B)(6)