FDA Adverse Event Injury Summary report: N

MEDTRONIC LEAD

MDR report key: 15960652 · Received December 12, 2022

Report

Report Number
2182208-2022-03967
Event Type
Injury
Date Received
December 12, 2022
Date of Event
April 15, 2022
Report Date
December 12, 2022
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. MEDTRONIC WAS MADE AWARE OF THIS EVENT THROUGH A SEARCH OF LITERATURE PUBLICATIONS. THIS EVENT OCCURRED OUTSIDE THE US. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. THE EVENT ONLY OCCURRED WITH ONE PATIENT BUT SPECIFIC DETAILS ON THE PATIENT WERE NOT PROVIDED. THE MODEL LISTED IN THE REPORT IS A REPRESENTATIVE OF THE MODEL FAMILY, AS THERE IS NO SPECIFIC MODEL LISTED. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. SINCE NO DEVICE ID WAS PROVIDED, IT IS UNKNOWN IF THIS EVENT HAS BEEN PREVIOUSLY REPORTED. A REQUEST FOR ADDITIONAL INFORMATION WILL BE MADE AND UPON RECEIPT A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY. REFERENCED ARTICLE: RAILROAD PATTERN OF ATRIAL INTERVALS IN A DUAL CHAMBER PACEMAKER PATIENT¿WHAT IS THE MECHANISM AND HOW TO MANAGE? THE JOURNAL OF INNOVATIONS IN CARDIAC RHYTHM MANAGEMENT. 2022; 13(4):4960¿4963. DOI: 10.19102/ICRM.2022.130406. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

A JOURNAL ARTICLE WAS REVIEWED THAT CONTAINED INFORMATION REGARDING THIS RIGHT ATRIAL (RA) LEAD. THE ARTICLE REPORTS A PATIENT WHOSE DEVICE RECORDED MANY ATRIAL HIGH-RATE (AHR) AND VENTRICULAR HIGH-RATE (VHR) EPISODES. THE ATRIAL INTERVALS SHOWED A RAILROAD PATTERN AND THE AHR EPISODES WERE DUE TO FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEAD WAS REPROGRAMMED AND APPEARS TO BE STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
386186 MEDTRONIC LEAD IMPLANTABLE PACEMAKER PULSE-GENERATOR DXY MEDTRONIC, INC. MDT-LEAD

Patients

Seq Age Sex Outcome Treatment
1 55 YR Female Required Intervention MDT IPG