FDA Adverse Event Injury Summary report: N

ISPAN PERFLUOROPROPANE (C3F8) GAS

MDR report key: 15957328 · Received December 12, 2022

Report

Report Number
2518435-2022-00048
Event Type
Injury
Date Received
December 12, 2022
Date of Event
January 13, 2022
Report Date
April 28, 2023
Manufacturer
AIRGAS THERAPEUTICS LLC
Product Code
LPO
UDI-DI
00380657971022
PMA / PMN Number
P900066
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS PROVIDED IN SECTIONS H.6 AND H.10. THERE WAS NO SAMPLE RETURNED FOR EVALUATION. AN ANALYSIS OF THE RETAINED (MANUFACTURING) SAMPLE FOR THESE LOTS, SHOWED THAT THE PRODUCT WAS PERFLUOROPROPANE (PFP) AND MET ALL RELEASE CRITERIA. NOTHING WAS FOUND THAT COULD BE ATTRIBUTED TO REPORTED EVENT. A CHECK OF THE BATCH PRODUCTION RECORD FOR THIS LOT, SHOWED NO UNUSUAL MANUFACTURING ISSUES. A CHECK OF CONFIRMED COMPLAINTS FOR PFP CAUSING SUBCONJUNCTIVAL HEMORRHAGE OR HYPHEMA SHOWED NO COMPLAINTS SINCE THE BEGINNING OF 2015. BASED UPON THE INFORMATION PROVIDED, THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. AT THIS TIME, THE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED. MANUFACTURER WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT RECEIVED PERFLUORO PROPANE GAS AND EXPERIENCED MILD CONJUNCTIVAL HEMORRHAGE, HYPHEMA IN THE LEFT EYE AND RECEIVED MEDICINE TREATMENT FOR THE EVENTS. THE PATIENT'S SYMPTOM WAS RESOLVED COMPLETELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2181514 ISPAN PERFLUOROPROPANE (C3F8) GAS INTRAOCULAR GAS LPO AIRGAS THERAPEUTICS LLC NA 034914 00380657971022

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention| O