FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 15950957 · Received December 9, 2022

Report

Report Number
1911916-2022-00716
Event Type
Malfunction
Date Received
December 9, 2022
Date of Event
October 28, 2022
Report Date
November 25, 2022
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
50382903096539
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IT WAS REPORTED ON THREE SEPARATE OCCASIONS THERE WAS PARTICULATE WHILE SAMPLING. TO AID IN THE INVESTIGATION, THREE FTIR REPORTS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE FTIR REPORTS REFER TO TESTING 15ML CONICAL TUBES CONTAINING HIC POOL MATERIAL. ONE REPORT MENTIONS ONE SAMPLE, ANOTHER REPORT MENTIONS TWO SAMPLES AND THE THIRD REPORT THREE SAMPLES. THE REPORTS INCLUDE PHOTOS TAKEN AT TWO HUNDRED AND FOUR HUNDRED MAGNIFICATIONS. AS WELL AS COMPARISON CHARTS OF THE FOREIGN MATTER VERSUS THE FTIR LIBRARY. THE CONCLUSION IS THE FOREIGN MATTER IS SILICONE, THOUGH THE REPORT DOES NOT MENTION THE PERCENT MATCH TO THE REFERENCE FROM THE LIBRARY. NO OTHER INFORMATION COULD BE OBTAINED FROM THE REPORTS PROVIDED AND NO INFORMATION WAS PROVIDED ON WHERE THE FOREIGN MATTER WAS IN THE BD SYRINGE. THE SYRINGES ARE MANUFACTURED WITH MEDICAL GRADE SILICONE THAT IS APPLIED TO THE INNER WALL SYRINGE BARREL AND RUBBER STOPPER. THIS ALLOWS FOR SMOOTHER MOVEMENT OF THE RUBBER STOPPER IN THE SYRINGE BARREL. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT NUMBER 2090548. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE FTIR REPORTS PROVIDED BY THE CUSTOMER THE DEFECT REPORTED COULD NOT BE CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND IN THE BD LUER-LOK¿ TIP SYRINGE DURING A SAMPLING EVENT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 3 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THIS WAS FOUND DURING ROUTINE SAMPLING OF A PROCESS POOL. LOOSE PARTICULATE FOUND WITHIN SINGLE USE SAMPLING VESSEL." "ON (B)(6) 2022 IN PA2S ROOM 512 DURING A SAMPLING EVENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2542914 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 2090548 50382903096539

Patients

Seq Age Sex Outcome Treatment
1 Unknown