FDA Adverse Event Malfunction Summary report: N

HEATER/COOLER 3T 110V/60 HZ

MDR report key: 15948376 · Received December 9, 2022

Report

Report Number
9611109-2022-00637
Event Type
Malfunction
Date Received
December 9, 2022
Date of Event
November 9, 2022
Report Date
November 9, 2022
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
PMA / PMN Number
K191402
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

H10: COMPLAINTS DATABASE ANALYSIS REVEALED THAT NO SIMILAR EVENT ON THIS DEVICE OCCURRED SINCE ITS INSTALLATION IN 2014. THE MOST LIKELY ROOT CAUSE OF THE REPORTED ISSUE IS ASSOCIATED TO MOTOR BEARING DAMAGED BY THE CORROSION, WHICH CONSEQUENTLY DOES NOT ALLOW THE MOTOR SHAFT TO ROTATE. POSSIBLE CAUSES ARE WATER INFILTRATION OR WATER FORMATION DUE TO CONDENSATION OR HIGH TEMPERATURES AND HIGH LEVEL OF HUMIDITY IN THE STATIONARY PHASE. CAUSES RELATED TO ENVIRONMENTAL CONDITIONS.

Additional Manufacturer Narrative · 0

LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN UNITED STATES. A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE THE DEVICE AND THE ERROR EE7 WAS CONFIRMED ON THE PATIENT SIDE. THE CIRCULATOR MOTOR (ITEM 96-410-719) AND EROSION KIT (ITEM 005-21-0061) WERE REPLACED. THE DEVICE WAS TESTED AND RETURNED TO SERVICE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A 3T HEATER-COOLER DEVICE DISPLAYED A TEMPERATUR PROBLEM ERROR (E07) DURING MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2701882 HEATER/COOLER 3T 110V/60 HZ CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-85

Patients

Seq Age Sex Outcome Treatment
1 Unknown