FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM

MDR report key: 15946123 · Received December 8, 2022

Report

Report Number
0002023141-2022-03057
Event Type
Injury
Date Received
December 8, 2022
Date of Event
November 18, 2022
Report Date
April 12, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019515
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-03058. ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS AND INFECTION EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS AND INFECTION HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMITIES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2022-03058-1. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. D9: DATE RETURNED TO MANUFACTURER WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE UPDATED: 3331. H10: NARRATIVE/DATA WAS UPDATED. DHR REVIEW WAS COMPLETED FOR THE ASSOCIATED SUBJECT LOT NUMBER 62365203. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHRS. LOTS WERE INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD REVIEW WAS PERFORMED FOR THE ASSOCIATED LOT NUMBER 62365203 (ST# 2292) AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE ASSOCIATED LOT NUMBERS 62365203 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATIONS, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATIONS, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENTS REMAIN NON-VERIFIABLE AS THEY ARE A MEDICAL CONDITION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT ON TOOTH SITES #14 AND #16 WERE REMOVED DUE TO BONE LOSS AND INFECTION.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION WAS RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2886074 IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB11 62365203 00889024019515

Patients

Seq Age Sex Outcome Treatment
1 73 YR Female Required Intervention