FDA Adverse Event Injury Summary report: N

AVISTA MRI

MDR report key: 15945477 · Received December 8, 2022

Report

Report Number
3006630150-2022-06872
Event Type
Injury
Date Received
December 8, 2022
Date of Event
May 1, 2022
Report Date
December 8, 2022
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
UDI-DI
08714729904823
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

EXACT DATE UNKNOWN, EVENT OCCURRED IN (B)(6) 2022. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS-MRI. UPN: M365SC2408740. MODEL: SC-2408-74. SERIAL: (B)(4). BATCH: 7072971.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT EXPERIENCED INADEQUATE STIMULATION DUE TO LEAD MIGRATION THAT WAS CONFIRMED VIA X-RAY. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEADS WILL NOT BE RETURNED PER HOSPITAL POLICY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448222 AVISTA MRI STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2408-74 5010485 08714729904823

Patients

Seq Age Sex Outcome Treatment
1 53 YR Female Required Intervention