FDA Adverse Event Malfunction Summary report: N

SYNCHRON® LX20 PRO CLINICAL SYSTEM

MDR report key: 1594515 · Received February 4, 2010

Report

Report Number
2050012-2010-00033
Event Type
Malfunction
Date Received
February 4, 2010
Date of Event
January 11, 2010
Report Date
February 4, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K965240
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

UPON TROUBLESHOOTING THE CUSTOMER DISCOVERED THE REAGENT SYRINGE WAS LOOSE WHERE IT CONNECTS TO THE T-VALVE. PER CUSTOMER, MAINTENANCE TO THESE PARTS HAD NOT BEEN PERFORMED RECENTLY. SERVICE REPLACED THE SYRINGE AND THE T-VALVE. UPON RECUR, THE ISSUE COULD AFFECT OTHER CHEMISTRIES INCLUDING POTENTIALLY PIVOTAL CHEMISTRIES, AND TREATMENT COULD BE INITIATED OR WITHHELD BASED UPON THE ERRONEOUS RESULT. A CLEAR ROOT CAUSE IS NOT KNOWN AT THIS TIME. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING A SUPPRESSED MAGNESIUM (MG) RESULT GENERATED BY THE SYNCHRON LX20 PRO CLINICAL SYSTEM. THE INITIAL "SUPPRESSED OIR LO" MG RESULT WAS INTERPRETED TO BE <0.1 MG/DL AND REPORTED OUT AS 0.1 MG/DL. THE SPECIMEN WAS TESTED ON A DIFFERENT INSTRUMENT AND OBTAINED RESULT WAS "SUPPRESSED OIR HI". THE SPECIMEN WAS DILUTED 1:2, AND THEN RETESTED AND MG RESULT WAS 7.8 MG/DL (INSTRUMENT UNKNOWN) THE PATIENT WAS REDRAWN AND THE SAMPLE WAS TESTED ON THE LX20 ANALYZER WHICH GENERATED A "SUPPRESSED OIR HI" RESULT. THE REDRAWN SAMPLE WAS DILUTED 1:3 AND RE-TESTED, AND A RESULT OF 8.1 MG/DL WAS GENERATED. (INSTRUMENT UNKNOWN) THE RESULT WAS AMENDED. THERE WAS NO EFFECT TO THE PATIENT. THE ERRONEOUS RESULT WAS FOR A LABORING PATIENT RECEIVING A MG DRIP, AND THE LOW RESULT WAS NOT BELIEVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® LX20 PRO CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. LX20 PRO

Patients

Seq Age Sex Outcome Treatment
1