FDA Adverse Event
Malfunction
Summary report: N
STATSENSOR CREATININE HOSPITAL METER
MDR report key: 1593542
·
Received January 27, 2010
Report
- Report Number
- 1219029-2010-00001
- Event Type
- Malfunction
- Date Received
- January 27, 2010
- Date of Event
- January 15, 2010
- Report Date
- January 18, 2010
- Manufacturer
- NOVA BIOMEDICAL
- Product Code
- CGL
- PMA / PMN Number
- K070068
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
CUSTOMER REPORTED THAT A PT CREATININE RESULT FROM A STATSENSOR METER WAS 1.1 MG/DL VERSUS A RESULT OF 3.0 MG/DL FROM A BECKMAN DXC ANALYZER. SAMPLE TYPE PLASMA FROM A GREEN TOP TUBE. QUALITY CONTROL RUN BEFORE AND AFTER THE EVENT WERE ACCEPTABLE. NO MAINTENANCE WAS DONE TO DEVICE BEFORE OR AFTER THE EVENT. AFTER THE REPORTED EVENT COMPARISON TESTING WAS PERFORMED BETWEEN THE STATSENSOR METER AND THE BECKMAN DXC ANALYZER WITHOUT INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STATSENSOR CREATININE HOSPITAL METER | CGL - ELECTRODE, ION BASED, ENZYMATIC CR | CGL | NOVA BIOMEDICAL | 44011 - METER | STRIP - 4909317249 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |