FDA Adverse Event Malfunction Summary report: N

EUREKA COLLIMATOR

MDR report key: 1593429 · Received January 22, 2010

Report

Report Number
1423380-2009-00031
Event Type
Malfunction
Date Received
January 22, 2010
Date of Event
December 29, 2009
Report Date
January 22, 2010
Manufacturer
MIDMARK CORPORATION
Product Code
IZW
PMA / PMN Number
K810598
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION OF THE UNIT REVEALED THAT THE TRIAC HAD FAILED. THE CERAMIC INSULATOR, WHICH IS USED TO REMOVE THE HEAT FROM THE TRIAC AND ELECTRICALLY ISOLATE THE UNIT FROM GROUND, WAS BROKEN, ISOLATING THE TRIAC FROM THE CHASSIS. THIS CAUSED THE TRIAC TO OVERHEAT AND FAIL IN A SHORTED MODE. THE TRIAC WAS REPLACED AND THE UNIT OPERATED NORMALLY. IN ADDITION, A COLLIMATOR (B) (4) WAS TESTED TO SIMULATE THE FAILURE MODE. THE CUSTOMER COMPLAINT ALLEGING "FIRE" ASSOCIATED WITH A SHORTED OUT LAMP CIRCUIT COULD NOT BE DUPLICATED.

Description of Event or Problem · 1

A DEALER SERVICE ENGINEER CALLED PROGENY ON (B) (6) 2009 REGARDING A COLLIMATOR, (B) (4), WHOSE LAMP CIRCUIT FAILED IN THE "ON" CONDITION CAUSING THE COVER TO BEGIN TO MELT. THE CUSTOMER ALLEGED THAT THE COLLIMATOR INTERNAL WIRING INSULATION CAUGHT FIRE AND CONTINUED TO BURN UNTIL THE WIRING CONNECTION FAILED WHICH STOPPED THE ELECTRIC CURRENT AND BURNING CEASED. NO INJURIES WERE REPORTED. THE UNIT WAS REPAIRED BY PROGENY TECHNICIAN ON 12/28/2009.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EUREKA COLLIMATOR IZW MIDMARK CORPORATION LINEAR II

Patients

Seq Age Sex Outcome Treatment
1