EASYSTAND EVOLV
Report
- Report Number
- 2183634-2010-00001
- Event Type
- Malfunction
- Date Received
- February 1, 2010
- Date of Event
- January 4, 2010
- Report Date
- January 26, 2010
- Manufacturer
- ALTIMATE MEDICAL, INC.
- Product Code
- INO
- PMA / PMN Number
- K062402
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
ADDITIONAL CATALOG NUMBER: PNG50024. JANUARY 27, 2010. EVAL SUMMARY: THE ORIGINAL EASYSTAND EVOLV GLIDER WAS RETURNED TO ALTIMATE MEDICAL ON JANUARY 25, 2010. UPON REVIEW, THE ACTUAL CAUSE OF WHY THE CASTER BROKE OFF THIS UNIT COULD NOT BE DETERMINED. IT CAN BE SPECULATED THAT THE FOLLOWING MAY HAVE OCCURRED WHICH CAUSED THIS TO HAPPEN: THE UNIT MAY HAVE BEEN DROPPED WHEN BEING DELIVERED WHICH CAUSED THE CASTER TO BEND AND WITH USE OVER TIME CAUSED THIS CASTER TO BREAK. BASED ON REVIEW OF THE EVOLV BASE, THE AREA WHERE THE CASTER THREADS INTO LOOKS TO BE DAMAGED AND THERE IS PAINT SCRAPED OFF OF THE BASE IN THAT AREA. THE CASTER MAY HAVE LOOSENED OVER TIME CAUSING THE CASTER TO BACK OUT OF THE BASE WHICH MAY HAVE CAUSED THIS TO OCCUR. THE CASTER MAY HAVE BEEN DAMAGED WHEN MOVING THE UNIT, EITHER BY RUNNING INTO A WALL OR BY GOING OVER A THRESHOLD. THE BASE SHOWING DAMAGE COULD ALSO BE A FACTOR IN THIS SCENARIO. WE HAVE BEEN IN CONTACT WITH THE CASTER SUPPLIER, AND THEY WILL BE REVIEWING THE CASTER AS WELL TO SEE IF THEY CAN PROVIDE US WITH ANY ADDITIONAL INFORMATION.
A USER WAS SITTING IN AN EASYSTAND EVOLV GLIDER ABOUT HALF WAY UP TO THE STANDING POSITION WHEN ONE OF THE REAR CASTERS BROKE OFF THE UNIT. THE USER'S ATTENDANT WAS NEXT TO HIM AND SUPPORTED THE EASYSTAND, THEREFORE, THE USER WAS NOT INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EASYSTAND EVOLV | 890.3110 | INO | ALTIMATE MEDICAL, INC. | EVOLV ADULT GLIDER | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Other |