HMS PLUS INSTRUMENT
Report
- Report Number
- 2184009-2022-00399
- Event Type
- Malfunction
- Date Received
- December 7, 2022
- Date of Event
- November 9, 2022
- Report Date
- February 2, 2023
- Manufacturer
- PERFUSION SYSTEMS
- Product Code
- JOX
- PMA / PMN Number
- K101271
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CONCLUSION: THE COMPLAINT WAS NOT CONFIRMED FOR THE HMS PLUS INSTRUMENT'S LIQUID QUALITY CONTROL (QC) NUMBERS BEING OFF. THE SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE OR FIND ANY ISSUES. THE SERVICE TECHNICIAN CLEANED THE LIFT WIRE AND COMPLETED A FULL DIAGNOSTIC CHECK AS A PRECAUTION. PREVENTIVE MAINTENANCE/TESTING WAS PERFORMED PER SPECIFICATIONS. NO PATIENT/CLINICAL SAFETY ISSUES WERE REPORTED. TRENDS FOR ISSUES WITH THIS PRODUCT ARE REVIEWED AT QUARTERLY QUALITY MEETINGS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
DEVICE EVALUATION:THE REPORTED LIQUID QUALITY CONTROL (QC) NUMBERS BEING OFF WAS NOT VERIFIED DURING SERVICE. THE SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE OR FIND ANY ISSUES. THE SERVICE TECHNICIAN CLEANED THE LIFT WIRE AND COMPLETED A FULL DIAGNOSTIC CHECK AS A PRECAUTION. POST-REPAIR TESTING WAS PERFORMED PER SPECIFICATIONS. NOTE: INSTRUMENT WAS NOT RETURNED TO MEDTRONIC FACILITY BUT WAS SERVICED BY FIELD SERVICE TECHNICIAN. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A HMS PLUS INSTRUMENT, IT WAS REPORTED THAT THE LIQUID QUALITY CONTROL (QC) NUMBERS WERE OFF. THE INSTRUMENT WAS USED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT WET CONTROLS ARE PERFORMED WEEKLY. NO ERROR WAS REPORTED. HEPARIN WAS ADMINISTERED BUT NOT BASED OFF THE ACT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1470194 | HMS PLUS INSTRUMENT | ANALYZER, HEPARIN, AUTOMATED | JOX | PERFUSION SYSTEMS | 30514 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |